Diagnostic performance evaluation of the Siemens Healthineers Atellica HEMA 580 hematology analyzer for abnormal cell detection and optimization of immature myeloid cell cut-offs
Clinical Chemistry and Laboratory Medicine (CCLM) · 24.06.2026 · 64 (7) · 1676-1686
DOI: 10.1515/cclm-2026-0083 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Clinical Chemistry and Laboratory Medicine (CCLM), online erschienen am 03.03.2026. Verfasst von Vincenzo De Iuliis und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Pathology, G. Mazzini Civil Hospital, Teramo, Italy. Hauptthemen laut MEDLINE: Myeloid Cells, Hematology. Die Arbeit nennt 24 Literaturangaben.
Abstract (Originalsprache)
Objectives
We evaluated the diagnostic performance of the Atellica HEMA 580 hematology analyzer for complete blood count (CBC) parameters compared with the ADVIA 2120i and assessed its capability to flag abnormal cells, and determined optimal cut-offs for immature myeloid cell detection.
Methods
A total of 400 K 2 EDTA blood samples were analyzed. Peripheral blood smears were independently reviewed by two clinical pathologists, each performing a 200-cell differential count. Analytical correlation, regression, Bland–Altman, and receiver operating characteristic (ROC) curve analyses were used to assess performance. Optimal cut-offs for immature myeloid cell (IMC) detection were determined using ROC analysis and the Youden index.
Results
Manual microscopy identified 23 samples with blasts (≥0.5 %), 177 with immature myeloid cells, and 61 with nucleated red blood cells (NRBCs). ROC analysis showed good accuracy for IMC detection using the IMG (immature granulocytes) parameter (AUC=0.81). An IMG cut-off >0.8 provided the best balance between sensitivity (40.7 %) and specificity (96.9 %), while large immature cells (LICs) >0.9 yielded higher sensitivity (60.5 %) at the expense of specificity (82.5 %). The Atellica HEMA 580 also demonstrated superior NRBC detection compared to ADVIA 2120i.
Conclusions
The Atellica HEMA 580 provided reliable performance in detecting immature myeloid cells and NRBCs. The IMG cut-off >0.8, determined through ROC optimization, supports its implementation for enhancing laboratory workflow efficiency while maintaining diagnostic accuracy. These findings support the clinical implementation of the Atellica HEMA 580 for routine hematology laboratories, particularly for enhanced detection of clinically significant immature myeloid populations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Clinical Pathology, G. Mazzini Civil Hospital, Teramo, Italy.
ORCID-Profile: Vincenzo De Iuliis, Sofia Chiatamone Ranieri
Zitieren
APA
De Iuliis, V., & Chiatamone Ranieri, S. (2026). Diagnostic performance evaluation of the Siemens Healthineers Atellica HEMA 580 hematology analyzer for abnormal cell detection and optimization of immature myeloid cell cut-offs. Clinical Chemistry and Laboratory Medicine (CCLM), 64(7), 1676-1686. https://doi.org/10.1515/cclm-2026-0083
Vancouver
De Iuliis V, Chiatamone Ranieri S. Diagnostic performance evaluation of the Siemens Healthineers Atellica HEMA 580 hematology analyzer for abnormal cell detection and optimization of immature myeloid cell cut-offs. Clinical Chemistry and Laboratory Medicine (CCLM). 2026;64(7):1676-1686. doi:10.1515/cclm-2026-0083
BibTeX
@article{DeIuliis2026,
author = {De Iuliis, Vincenzo and Chiatamone Ranieri, Sofia},
title = {Diagnostic performance evaluation of the Siemens Healthineers Atellica HEMA 580 hematology analyzer for abnormal cell detection and optimization of immature myeloid cell cut-offs},
journal = {Clinical Chemistry and Laboratory Medicine (CCLM)},
year = {2026},
volume = {64},
number = {7},
pages = {1676--1686},
doi = {10.1515/cclm-2026-0083}
}
EUVIMED