Computer-based analysis of the taper connection strength of different revision head and adapter sleeve designs

Johanna Meier iD, Jessica Hembus iD, Rainer Bader, Danny Vogel iD

Biomedical Engineering / Biomedizinische Technik · 2024

DOI: 10.1515/bmt-2023-0066 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Engineering / Biomedizinische Technik, 2024. Verfasst von Johanna Meier und 3 weiteren Autor:innen. Beteiligte Einrichtung: Biomechanics and Implant Technology Research Laboratory, Department of Orthopaedics, Rostock University Medical Center, Rostock, Germany. Hauptthemen laut MEDLINE: Hip Prosthesis, Arthroplasty, Replacement, Hip. Die Arbeit nennt 35 Literaturangaben. Gefördert von: Bundesministerium für Bildung und Forschung.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Ceramic revision heads, equipped with titanium adapter sleeves, are used in femoral head revision in total hip arthroplasty to avoid ceramic fracture due to the damaged taper.

Methods

A finite element analysis of the taper connection strength of revision heads with varying head diameters combined with adapter sleeves of different lengths was conducted. The influence of various assembly forces, head diameter, and length of the adapter sleeves was evaluated. For two combinations, the pattern of contact pressure was evaluated when applying a simplified joint load (3 kN, 45° load angle). Experimental validation was conducted with 36 mm heads and adapter sleeves in size S, as well as 28 mm heads and adapter sleeves in size XL.

Results

The pull-off force increased with higher assembly forces. Using larger head diameters and adapter sleeves led to decreased pull-off forces, a reduced contact surface, and less contact pressure. The contact pressure showed significant peaks and a diagonal pattern under 45° angle loading when assembly forces were less than 4 kN, and larger adapter sleeves were utilized.

Conclusion

A sufficient assembly force should be ensured intraoperatively, especially with an increasing head diameter and adapter sleeve size, as lower assembly forces might lead to reduced taper connection strength.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hip ProsthesisArthroplasty, Replacement, HipHumansProsthesis DesignSoftwareFemur HeadComputersProsthesis Failure

Stichwörter der Autor:innen

adapter sleevesfinite element analysisrevision headtaper connectiontotal hip arthroplasty

Autor:innen und Institutionen

  • Biomechanics and Implant Technology Research Laboratory, Department of Orthopaedics, Rostock University Medical Center, Rostock, Germany.

ORCID-Profile: Johanna Meier, Jessica Hembus, Danny Vogel

Förderung

  • Bundesministerium für Bildung und Forschung

Zitieren

APA

Meier, J., Hembus, J., Bader, R., & Vogel, D. (2024). Computer-based analysis of the taper connection strength of different revision head and adapter sleeve designs. Biomedical Engineering / Biomedizinische Technik. https://doi.org/10.1515/bmt-2023-0066

Vancouver

Meier J, Hembus J, Bader R, Vogel D. Computer-based analysis of the taper connection strength of different revision head and adapter sleeve designs. Biomedical Engineering / Biomedizinische Technik. 2024. doi:10.1515/bmt-2023-0066

BibTeX
@article{Meier2024,
  author = {Meier, Johanna and Hembus, Jessica and Bader, Rainer and Vogel, Danny},
  title = {Computer-based analysis of the taper connection strength of different revision head and adapter sleeve designs},
  journal = {Biomedical Engineering / Biomedizinische Technik},
  year = {2024},
  doi = {10.1515/bmt-2023-0066}
}
Kostenlos starten →