A comparison of resting metabolic rate measurement in adults, using a ventilated canopy or a mixing‐chamber system with a silicone mask

Yair Pincu iD, F N U Sakshi, Grant A Chesbro, Akram Falahati, Rebecca D Larson

Physiological Reports · 2026-03 · 14 (6)

DOI: 10.14814/phy2.70837 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Physiological Reports, online erschienen am 25.03.2026. Verfasst von Yair Pincu und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Health and Exercise Science, University of Oklahoma, Norman, Oklahoma, USA. Hauptthemen laut MEDLINE: Basal Metabolism, Calorimetry, Indirect, Masks. Die Arbeit nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The most common method for measuring resting metabolic rate (RMR) via indirect calorimetry (IC) is using a ventilated canopy system (canopy). Yet, many use a facemask (mask). The purpose of this study was to compare RMR using a canopy and a mask. RMR was measured in 26 participants using a canopy or a mask, followed immediately by the opposite mode in a randomized fashion. RMR was compared between the canopy and mask in males and females using a 2 × 2 repeated‐measures ANOVA. Agreement between the 2 modes and typical error of measurement (TEM) were calculated. Data are mean ± SD. Resting VO 2 and RER were significantly different (VO 2 = 3.72 ± 0.70 vs. 3.46 ± 0.56 mL/kg/min; RER = 0.75 ± 0.04 vs. 0.80 ± 0.04 in canopy vs. mask, p < 0.05). RMR was lower using the mask (1719.71 ± 377.43 vs. 1791.23 ± 387.61 kcal/day, p = 0.026). Mask and canopy RMR showed poor agreement (95% Limit of agreement = −363.5 to 220.5 kcal/day). TEM was larger than the bias between the 2 methods (105.33 kcal/day), which suggests significant random error. Males ( n = 14) had a larger and females ( n = 12) smaller difference in RMR between canopy and mask (−79.14 ± 164.27, p = 0.063, vs. −62.62 ± 135.160 kcal/day, p = 0.166, respectively). We conclude that non‐ventilated facemask cannot produce comparable metabolic data and should not be used for RMR in place of a ventilated canopy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Basal MetabolismCalorimetry, IndirectMasksHumansMaleFemaleAdultOxygen Consumption

Stichwörter der Autor:innen

indirect calorimetrymetabolic carttotal daily energy expenditureventilated hood

Autor:innen und Institutionen

  • Department of Health and Exercise Science, University of Oklahoma, Norman, Oklahoma, USA.
  • Harold Hamm Diabetes Center, University of Oklahoma Health Campus, Oklahoma City, Oklahoma, USA.
  • TSET Health Promotion Research Center, The University of Oklahoma Health Campus, Oklahoma City, Oklahoma, USA.
  • Program for Cellular and Behavioral Neurobiology, University of Oklahoma, Norman, Oklahoma, USA.

ORCID-Profile: Yair Pincu

Zitieren

APA

Pincu, Y., Sakshi, F. N. U., Chesbro, G. A., Falahati, A., & Larson, R. D. (2026). A comparison of resting metabolic rate measurement in adults, using a ventilated canopy or a mixing‐chamber system with a silicone mask. Physiological Reports, 14(6). https://doi.org/10.14814/phy2.70837

Vancouver

Pincu Y, Sakshi FNU, Chesbro GA, Falahati A, Larson RD. A comparison of resting metabolic rate measurement in adults, using a ventilated canopy or a mixing‐chamber system with a silicone mask. Physiological Reports. 2026;14(6). doi:10.14814/phy2.70837

BibTeX
@article{Pincu2026,
  author = {Pincu, Yair and Sakshi, F N U and Chesbro, Grant A and Falahati, Akram and Larson, Rebecca D},
  title = {A comparison of resting metabolic rate measurement in adults, using a ventilated canopy or a mixing‐chamber system with a silicone mask},
  journal = {Physiological Reports},
  year = {2026},
  volume = {14},
  number = {6},
  doi = {10.14814/phy2.70837}
}
Kostenlos starten →