Validity and reproducibility of a whole‐room indirect calorimeter for the estimation of VO 2 , VCO 2 , and resting metabolic rate
Physiological Reports · 2023
DOI: 10.14814/phy2.15658 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physiological Reports, 2023. Verfasst von Hege Berg Henriksen und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nutrition, Faculty of Medicine, Institute of Basic Medical Sciences, University of Oslo, Oslo, Norway. Hauptthemen laut MEDLINE: Basal Metabolism, Energy Metabolism. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Whole‐room indirect calorimeters (WRICs) provide accurate instruments for the measurement of respiratory exchange, energy expenditure, and macronutrient oxidation. Here, we aimed to determine the validity and reproducibility of a 7500 L WRIC for the measurement of ventilation rates and resting metabolic rate (RMR). Technical validation was performed with propane combustion tests ( n = 10) whereas biological reproducibility was tested in healthy subjects (13 women, 6 men, mean ± SD age 39.6 ± 15.3) in two 60 min measurements separated by 24 h. Subjects followed a run‐in protocol prior to measurements. The coefficient of variation (CV) and intraclass correlation coefficient (ICC) were calculated for ventilation rates of O2 (VO2), CO2 (VCO2), the respiratory quotient (RQ; VCO2/VO2), and RMR. Technical validation showed good validity with CVs ranging from 0.67% for VO 2 to 1.00% for energy expenditure. For biological reproducibility, CVs were 2.89% for VO 2 ; 2.67% for VCO 2 ; 1.95% for RQ; and 2.68% for RMR. With the exception of RQ (74%), ICCs were excellent for VO 2 (94%), VCO 2 (96%) and RMR (95%). Excluding participants that deviated from the run‐in protocol did not alter results. In conclusion, the 7500 L WRIC is technically valid and reproducible for ventilation rates and RMR.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Nutrition, Faculty of Medicine, Institute of Basic Medical Sciences, University of Oslo, Oslo, Norway.
- D & S Consulting Services, Inc., New York, New York, USA.
- Institute of Human Nutrition and Food Science, Christian-Albrechts University, Kiel, Germany.
- Division of Cancer Medicine, Department of Clinical Service, Oslo University Hospital, Oslo, Norway.
ORCID-Profile: Thomas Olsen
Zitieren
APA
Henriksen, H. B., Henriksen, C., de Sousa, A. R. S., Alavi, D. T., Augestad, E. M. S., Rising, R., Dörner, R., Hägele, F. A., Bosy-Westphal, A., Blomhoff, R., Ulven, S. M., & Olsen, T. (2023). Validity and reproducibility of a whole‐room indirect calorimeter for the estimation of VO 2 , VCO 2 , and resting metabolic rate. Physiological Reports. https://doi.org/10.14814/phy2.15658
Vancouver
Henriksen HB, Henriksen C, de Sousa ARS, Alavi DT, Augestad EMS, Rising R, et al. Validity and reproducibility of a whole‐room indirect calorimeter for the estimation of VO 2 , VCO 2 , and resting metabolic rate. Physiological Reports. 2023. doi:10.14814/phy2.15658
BibTeX
@article{Henriksen2023,
author = {Henriksen, Hege Berg and Henriksen, Christine and de Sousa, Ana Rita Sequeira and Alavi, Dena Treider and Augestad, Elin Maria Sandstad and Rising, Russell and Dörner, Rebecca and Hägele, Franziska Anna and Bosy-Westphal, Anja and Blomhoff, Rune and Ulven, Stine Marie and Olsen, Thomas},
title = {Validity and reproducibility of a whole‐room indirect calorimeter for the estimation of VO 2 , VCO 2 , and resting metabolic rate},
journal = {Physiological Reports},
year = {2023},
doi = {10.14814/phy2.15658}
}
EUVIMED