Repeatability of corneal measurements from the Casia2 anterior segment tomographer in a cataractous population
PLOS One · 2025
DOI: 10.1371/journal.pone.0328894 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS One, 2025. Verfasst von Achim Langenbucher und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Experimental Ophthalmology, Saarland University, Homburg/Saar, Germany. Hauptthemen laut MEDLINE: Cornea, Cataract, Tomography, Anterior Eye Segment. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
The purpose of this study was to investigate the repeatability of corneal power and thickness measures in a large patient cohort.
Methods
In this retrospective non-randomised cross sectional single-centre study we evaluated a dataset containing 900 Casia2 anterior segment tomography measurements from 300 eyes from 300 patients (3 repeat measurements each) taken prior to cataract surgery. Only complete measurements marked as ‘Successful’ and with a sequence of 3 measurements for each eye performed on the same day were considered. Keratometry (K), corneal back surface (P) and RealPower (R) were each decomposed into power vector components SEQ/C0/C45. The mean (MEAN) and standard deviations (SD) for each sequence of measurements were derived and analysed.
Results
For K the MEAN and SD SEQ/C0/C45 were 43.77/0.17/-0.02 D and 0.12/0.18/0.18 D, for P MEAN and SD were −6.13/-0.25/0.00 D and 0.01/0.03/0.02 D, and for R MEAN and SD were 42.77/-0.06/-0.03 D and 0.13/0.19/0.19 D. For K and R 20%/42%/41% and 24%/45%/44% of repeat measurements respectively fell outside limits of ±1/8 D. The cornea was slightly thicker at the apex than at the thinnest point (544.6 vs. 536.2 µm) and SD was 4.3 vs. 3.9 µm. Keratometry underestimates against-the-rule corneal astigmatism by around 0.23 D.
Conclusions
The variation of corneal power with repeat measurements is in a range of ±1/8 D for keratometric and RealPower SEQ. Where the Casia2 is used for lens power calculation, repeat measurements could help to improve the refractive outcome.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Experimental Ophthalmology, Saarland University, Homburg/Saar, Germany.
- Dr. Rolf M. Schwiete Center for Limbal Stem Cell and Aniridia Research, Saarland University, Homburg/Saar, Germany.
- Department of Ophthalmology, Semmelweis-University, Budapest, Hungary.
- School of Physical Sciences, The Open University, Milton Keynes, United Kingdom.
- Department of Ophthalmology and Optometry, Vienna University Hospital, Vienna, Austria.
- Department of Ophthalmology, Ludwig Maximilian University (LMU), Munich, Germany.
ORCID-Profile: Achim Langenbucher, Alan Cayless
Zitieren
APA
Langenbucher, A., Szentmáry, N., Cayless, A., Schartmüller, D., Lisy, M., Menapace, R., Wendelstein, J., & Leydolt, C. (2025). Repeatability of corneal measurements from the Casia2 anterior segment tomographer in a cataractous population. PLOS One. https://doi.org/10.1371/journal.pone.0328894
Vancouver
Langenbucher A, Szentmáry N, Cayless A, Schartmüller D, Lisy M, Menapace R, et al. Repeatability of corneal measurements from the Casia2 anterior segment tomographer in a cataractous population. PLOS One. 2025. doi:10.1371/journal.pone.0328894
BibTeX
@article{Langenbucher2025,
author = {Langenbucher, Achim and Szentmáry, Nóra and Cayless, Alan and Schartmüller, Daniel and Lisy, Marcus and Menapace, Rupert and Wendelstein, Jascha and Leydolt, Christina},
title = {Repeatability of corneal measurements from the Casia2 anterior segment tomographer in a cataractous population},
journal = {PLOS One},
year = {2025},
doi = {10.1371/journal.pone.0328894}
}
EUVIMED