Repeatability of corneal curvature with the MS-39 anterior segment tomographer in a cataractous population

Achim Langenbucher iD, Nóra Szentmáry, Alan Cayless, Muntadher Al Karam, Theo G Seiler, Jascha Wendelstein

PLOS One · 2025

DOI: 10.1371/journal.pone.0327863 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS One, 2025. Verfasst von Achim Langenbucher und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Experimental Ophthalmology, Saarland University, Homburg/Saar, Germany. Hauptthemen laut MEDLINE: Tomography, Optical Coherence, Cataract, Cornea. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The purpose of this study was to investigate the repeatability of optical coherence tomography in measuring radius of curvature (ROC) and asphericity (Q) of corneal epithelium, stroma and endothelium in a large patient cohort.

Methods

In this retrospective non-randomized cross sectional single-center study we evaluated a dataset containing 600 anterior segment optical coherence tomography measurements from 200 eyes (3 repeat measurements each) taken prior to cataract surgery. Measurements were exported containing height map data for the epithelium, stroma and endothelium surface. Model surfaces (floating sphere in a 3 (BFS3) and 6 mm zone (BFS6) and conic surface in a 3 (Conoid3) and 6 mm zone (Conoid6)) were fitted using nonlinear iterative optimization minimizing the height difference between measurement and model. The mean (MEAN) and standard deviations (SD) for each sequence of measurements were derived and analyzed.

Results

MEAN ROC for epithelium was 7.68/7.72/7.67/7.67 mm with BFS3/BFS6/Conoid3/Conoid6 and Q was −0.16/-0.19 with Conoid3/Conoid6. Corresponding values for stroma were 7.62/7.64/7.55/7.58 mm and −0.38/-0.24, and for endothelium they were 6.43/6.49/6.40/6.37 mm and −0.29/-0.33. Variations of the 3 repeat measurement ranged between 7–19/18–31/21–32 micron for epithelium/stroma/endothelium, and variation in Q was 0.14/0.20/0.14 with Conoid3 and 0.03/0.05/0.06 with Conoid6.

Conclusions

Fitting floating sphere or conoid model surfaces to the map data seems to be a very robust and reliable method for extracting ROC and asphericity from all data within a region of interest. For extraction of asphericity, the 6 mm zone should be used due to fewer variations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, Optical CoherenceCataractCorneaHumansFemaleMaleRetrospective StudiesAgedCross-Sectional StudiesMiddle AgedReproducibility of ResultsAged, 80 and overCataract ExtractionEpithelium, Corneal

Autor:innen und Institutionen

  • Department of Experimental Ophthalmology, Saarland University, Homburg/Saar, Germany.
  • Dr. Rolf M. Schwiete Center for Limbal Stem Cell and Aniridia Research, Saarland University, Homburg/Saar, Germany.
  • Department of Ophthalmology, Semmelweis-University, Budapest, Hungary.
  • School of Physical Sciences, The Open University, Milton Keynes, United Kingdom.
  • Institute for Refractive- and Ophthalmic Surgery (IROC), Zurich, Switzerland.
  • Department of Ophthalmology, Universitätsspital Zürich, Zurich, Switzerland.

ORCID-Profile: Achim Langenbucher

Zitieren

APA

Langenbucher, A., Szentmáry, N., Cayless, A., Al Karam, M., Seiler, T. G., & Wendelstein, J. (2025). Repeatability of corneal curvature with the MS-39 anterior segment tomographer in a cataractous population. PLOS One. https://doi.org/10.1371/journal.pone.0327863

Vancouver

Langenbucher A, Szentmáry N, Cayless A, Al Karam M, Seiler TG, Wendelstein J. Repeatability of corneal curvature with the MS-39 anterior segment tomographer in a cataractous population. PLOS One. 2025. doi:10.1371/journal.pone.0327863

BibTeX
@article{Langenbucher2025,
  author = {Langenbucher, Achim and Szentmáry, Nóra and Cayless, Alan and Al Karam, Muntadher and Seiler, Theo G and Wendelstein, Jascha},
  title = {Repeatability of corneal curvature with the MS-39 anterior segment tomographer in a cataractous population},
  journal = {PLOS One},
  year = {2025},
  doi = {10.1371/journal.pone.0327863}
}
Kostenlos starten →