Effect of autoclaving process on trueness and qualitative marginal fit of intraoral scan body manufactured from two materials: An in vitro study
PLOS One · 2025
DOI: 10.1371/journal.pone.0325068 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS One, 2025. Verfasst von Fábio Henrique de Paulo Costa Santos und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Occlusion, Fixed Prosthodontics, and Dental Materials, School of Dentistry, Federal University of Uberlandia, Uberlandia, Minas Gerais,…. Hauptthemen laut MEDLINE: Titanium, Dental Implants, Ketones, Dental Marginal Adaptation. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: CNPq, CAPES (Foundation Coordination for the Improvement of Higher Education Personnel), INCT Oral Health and Dentistry, FAPEMIG.
Abstract (Originalsprache)
Statement of the problem It is not clear the effect of multiple autoclaving processes on the accuracy of scan bodies. Autoclaving dental disposables enables reuse, reducing the environmental impact of raw materials and waste.
Purpose
This in vitro laboratory study evaluated the effect of the autoclaving process on the intraoral scan body (SB) manufactured from different materials (Titanium and PEEK), positioned at the implant and abutment levels.
Methods
Two models, each with 10 implants, were created to evaluate scan body trueness made of two materials (Titanium and PEEK), and implant junction positioned at two levels (implant and abutment levels). Analyses were conducted at five distinct time points: T0 (control group with new SB), T1 (after one autoclaving cycle), T2 (10 cycles), T3 (50 cycles), and T4 (100 cycles). Forty SBs were divided into four groups (n = 10), fixed on the models according to the initial positioning, using the recommended torque, and scanned 10 times using an intraoral Sccaner (InEOS X5 scanner, Dentsply Sirona). The surface deviation was evaluated by .STL mesh overlay using an image software (Geomagic Control X software). Microscopic analyses of marginal adaptation were performed simultaneously with the digital analysis using an optical microscope (Mitutoyo™-500 Optical Microscope) with 40x magnification at four places (mesial, distal, buccal, and lingual). Statistical analysis was performed using statistical software (SPSS 29.0 software), with α = 0,05. Data were subjected to the Shapiro-Wilk normality test. To assess spatial variation after the cycles, Repeated Measures ANOVA with Bonferroni correction was used. For microscopic qualitative analysis, percentile was used to check the prevalence of clinically adapted and misfitting faces.
Results
Statistically significant differences in mesh deformation were detected, with PEEK SB showing greater deformation compared to Titanium, especially at the abutment level, after 100 cycles (P < 0.01). Microscopic analysis revealed that 100% of the faces were classified as “Clinically Adapted” by the three calibrated evaluators (KAPPA > 0.08).
Conclusions
Titanium scan bodies showed less surface deviation than PEEK but both remained below 50 µm. Internal-junction scan bodies were more affected than abutment-level ones. All scan bodies stayed within acceptable limits and viable after 100 autoclaving cycles.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Department of Occlusion, Fixed Prosthodontics, and Dental Materials, School of Dentistry, Federal University of Uberlandia, Uberlandia, Minas Gerais, Brazil.
- Department of General Practice, School of Dentistry, Virginia Commonwealth University, Richmond, Virginia, United States of America.
ORCID-Profile: Patrícia Santos de Melo, Karla Zancope
Förderung
- CNPq
- CAPES (Foundation Coordination for the Improvement of Higher Education Personnel), INCT Oral Health and Dentistry
- FAPEMIG
Zitieren
APA
Costa Santos, F. H. P., de Melo, P. S., Borella, P. S., Neves, F. D. D., & Zancope, K. (2025). Effect of autoclaving process on trueness and qualitative marginal fit of intraoral scan body manufactured from two materials: An in vitro study. PLOS One. https://doi.org/10.1371/journal.pone.0325068
Vancouver
Costa Santos FHP, de Melo PS, Borella PS, Neves FDD, Zancope K. Effect of autoclaving process on trueness and qualitative marginal fit of intraoral scan body manufactured from two materials: An in vitro study. PLOS One. 2025. doi:10.1371/journal.pone.0325068
BibTeX
@article{CostaSantos2025,
author = {Costa Santos, Fábio Henrique de Paulo and de Melo, Patrícia Santos and Borella, Paulo Sérgio and Neves, Flávio Domingues das and Zancope, Karla},
title = {Effect of autoclaving process on trueness and qualitative marginal fit of intraoral scan body manufactured from two materials: An in vitro study},
journal = {PLOS One},
year = {2025},
doi = {10.1371/journal.pone.0325068}
}