EUVIMEDEUVIMED entdecken
Open Access · CC0

Morphometric characterization and temporal temperature measurements during hepatic microwave ablation in swine

Nicole A Varble iD, Ivane Bakhutashvili, Sheridan L Reed, Jose Delgado, Zoi Tokoutsi iD, Bruno Frackowiak iD, Marco Baragona, John W Karanian, Bradford J Wood, William F Pritchard iD

PLOS ONE · 2023

DOI: 10.1371/journal.pone.0289674 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS ONE, 2023. Verfasst von Nicole A Varble und 9 weiteren Autor:innen. Beteiligte Einrichtung: Center for Interventional Oncology, Radiology and Imaging Sciences, Clinical Center, National, Institutes of Health, Bethesda, Maryland, United State…. Hauptthemen laut MEDLINE: Liver. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC0). Gefördert von: NIH Clinical Center, HORIZON EUROPE Marie Sklodowska-Curie Actions, Philips.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Heat-induced destruction of cancer cells via microwave ablation (MWA) is emerging as a viable treatment of primary and metastatic liver cancer. Prediction of the impacted zone where cell death occurs, especially in the presence of vasculature, is challenging but may be achieved via biophysical modeling. To advance and characterize thermal MWA for focal cancer treatment, an in vivo method and experimental dataset were created for assessment of biophysical models designed to dynamically predict ablation zone parameters, given the delivery device, power, location, and proximity to vessels. Materials and methods MWA zone size, shape, and temperature were characterized and monitored in the absence of perfusion in ex vivo liver and a tissue-mimicking thermochromic phantom (TMTCP) at two power settings. Temperature was monitored over time using implanted thermocouples with their locations defined by CT. TMTCPs were used to identify the location of the ablation zone relative to the probe. In 6 swine, contrast-enhanced CTs were additionally acquired to visualize vasculature and absence of perfusion along with corresponding post-mortem gross pathology.

Results

Bench studies demonstrated average ablation zone sizes of 4.13±1.56cm 2 and 8.51±3.92cm 2 , solidity of 0.96±0.06 and 0.99±0.01, ablations centered 3.75cm and 3.5cm proximal to the probe tip, and temperatures of 50 ºC at 14.5±13.4s and 2.5±2.1s for 40W and 90W ablations, respectively. In vivo imaging showed average volumes of 9.8±4.8cm 3 and 33.2±28.4cm 3 and 3D solidity of 0.87±0.02 and 0.75±0.15, and gross pathology showed a hemorrhagic halo area of 3.1±1.2cm 2 and 9.1±3.0cm 2 for 40W and 90W ablations, respectfully. Temperatures reached 50ºC at 19.5±9.2s and 13.0±8.3s for 40W and 90W ablations, respectively.

Conclusion

MWA results are challenging to predict and are more variable than manufacturer-provided and bench predictions due to vascular stasis, heat-induced tissue changes, and probe operating conditions. Accurate prediction of MWA zones and temperature in vivo requires comprehensive thermal validation sets.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LiverAnimalsSwineMicrowavesTemperatureAblation TechniquesRadiofrequency Ablation

Autor:innen und Institutionen

  • Center for Interventional Oncology, Radiology and Imaging Sciences, Clinical Center, National, Institutes of Health, Bethesda, Maryland, United States of America.
  • Philips, Best, The Netherlands.
  • Fischell Department of Bioengineering, University of Maryland, College Park, Maryland, United States of America.
  • Bioengineering and National Cancer Institute Center, Bethesda, Maryland, United States of America.

ORCID-Profile: Nicole A Varble, Zoi Tokoutsi, Bruno Frackowiak, William F Pritchard

Förderung

  • NIH Clinical Center
  • HORIZON EUROPE Marie Sklodowska-Curie Actions
  • Philips
  • National Institutes of Health

Zitieren

APA

Varble, N. A., Bakhutashvili, I., Reed, S. L., Delgado, J., Tokoutsi, Z., Frackowiak, B., Baragona, M., Karanian, J. W., Wood, B. J., & Pritchard, W. F. (2023). Morphometric characterization and temporal temperature measurements during hepatic microwave ablation in swine. PLOS ONE. https://doi.org/10.1371/journal.pone.0289674

Vancouver

Varble NA, Bakhutashvili I, Reed SL, Delgado J, Tokoutsi Z, Frackowiak B, et al. Morphometric characterization and temporal temperature measurements during hepatic microwave ablation in swine. PLOS ONE. 2023. doi:10.1371/journal.pone.0289674

BibTeX
@article{Varble2023,
  author = {Varble, Nicole A and Bakhutashvili, Ivane and Reed, Sheridan L and Delgado, Jose and Tokoutsi, Zoi and Frackowiak, Bruno and Baragona, Marco and Karanian, John W and Wood, Bradford J and Pritchard, William F},
  title = {Morphometric characterization and temporal temperature measurements during hepatic microwave ablation in swine},
  journal = {PLOS ONE},
  year = {2023},
  doi = {10.1371/journal.pone.0289674}
}
Kostenlos starten →