Scatterer spacing based on Gabor atoms matched from harmonic ultrasound echoes to improve assessment of microwave‐induced thermal lesions

Xiuhua Zeng, Yufeng Zhang, Zhiyao Li, Jian Yang, Lian Gao, Junhua Zhang

Medical Physics · 2024

DOI: 10.1002/mp.17384 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Xiuhua Zeng und 5 weiteren Autor:innen. Beteiligte Einrichtung: Engineering Research Center of Intelligent System and Advance Materials of Yunnan Province Universities, College of Physics & Electronic Engineering,…. Hauptthemen laut MEDLINE: Microwaves, Liver, Ultrasonography. Die Arbeit nennt 44 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Applied basic Research Foundation of Yunnan Province, Qianke-He-Foundation-ZK[2021]-General-300.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Microwave ablation (MWA) is a minimally invasive alternative for the treatment of unresectable liver tumors. To verify the effectiveness and safety of MWA, it is critical to measure the temperature variation and assess the regions of the microwave‐induced thermal lesions.

Purpose

Recent studies have indicated that the locations of optimally matched Gabor atoms (LOMGA) from ultrasound radiofrequency (RF) echo signals allow accurate and stable scatterer spacing estimation. Herein, a harmonic‐based LOMGA method is proposed to estimate the scatterer spacing for improving the assessment of microwave‐induced thermal lesions.

Methods

The mean scatterer spacing (MSS) is estimated via the LOMGA method incorporating the selection of concise atoms from separated second‐harmonic RF echo signals with the pulse‐inversion algorithm for thermal lesion evaluation. In vitro experiments, 10 fresh porcine liver samples were ablated at different time nodes during the ablation period, and 200 sets of second‐harmonic and fundamental RF echo signals were randomly selected from the regions of interest in the coagulated liver samples for MSS estimation. The means and standard deviations of the MSSs, as well as the linear regression for the mean MSSs, were calculated from fundamental and second‐harmonic signals for comparison and evaluation, the receiver operating characteristic (ROC) curves for the 200 sets of fundamental‐based and harmonic‐based MSS estimates from the 10 liver samples at five pairs of adjacent time nodes were calculated, and one‐way analysis of variance (ANOVA) tests were performed for the five pairs of adjacent time nodes. The fundamental and harmonic‐based p ‐values in the ANOVA tests and the areas under the ROC curves (AUCs) were calculated to statistically analyze the differences in the MSSs between adjacent time nodes.

Results

The harmonic‐based increments in the intensity variation and coherent components were larger than the fundamental‐based increments with the increasing ablation time. The harmonic‐based MSSs from the 10 liver samples at five pairs of adjacent time nodes were found to be highly statistically significant ( p < 0.01). Thus, the harmonic‐based MSSs had greater variations. Compared with the fundamental‐based results, for the five preset ST values, the average increment in the harmonic‐based mean slopes was 69.22% and the average decrement in the mean standard deviations was 11.67% for the linear‐fitting MSS results, and the results were statistically significant ( p < 0.05).

Conclusion

Harmonic‐based MSSs are more sensitive and robust to variations in coagulated tissues, which is advantageous for the assessment of microwave‐induced thermal lesions.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

MicrowavesLiverUltrasonographyAnimalsSwineScattering, RadiationAblation TechniquesAlgorithmsROC CurveTemperature

Stichwörter der Autor:innen

Gabor atomharmonic RF echo signalmicrowave ablationscatterer spacingthermal lesion

Autor:innen und Institutionen

  • Engineering Research Center of Intelligent System and Advance Materials of Yunnan Province Universities, College of Physics & Electronic Engineering, Qujing Normal University, Qujing, Yunnan, China.
  • Department of Electronic Engineering, Information School, Yunnan University, Kunming, Yunnan, China.
  • The Third Affiliated Hospital of Kunming Medical University, Kunming, Yunnan, China.

Förderung

  • National Natural Science Foundation of China
  • Applied basic Research Foundation of Yunnan Province
  • Qianke-He-Foundation-ZK[2021]-General-300

Zitieren

APA

Zeng, X., Zhang, Y., Li, Z., Yang, J., Gao, L., & Zhang, J. (2024). Scatterer spacing based on Gabor atoms matched from harmonic ultrasound echoes to improve assessment of microwave‐induced thermal lesions. Medical Physics. https://doi.org/10.1002/mp.17384

Vancouver

Zeng X, Zhang Y, Li Z, Yang J, Gao L, Zhang J. Scatterer spacing based on Gabor atoms matched from harmonic ultrasound echoes to improve assessment of microwave‐induced thermal lesions. Medical Physics. 2024. doi:10.1002/mp.17384

BibTeX
@article{Zeng2024,
  author = {Zeng, Xiuhua and Zhang, Yufeng and Li, Zhiyao and Yang, Jian and Gao, Lian and Zhang, Junhua},
  title = {Scatterer spacing based on Gabor atoms matched from harmonic ultrasound echoes to improve assessment of microwave‐induced thermal lesions},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17384}
}
Kostenlos starten →