Fachartikel in der Zeitschrift F1000Research, online erschienen am 22.09.2023. Verfasst von Anush Nayak und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Allied Health Sciences,Faculty of Life & Allied Health Sciences(FLAHS), Ramaiah University of Applied Sciences, Bangalore, Karnataka, 5…. Hauptthemen laut MEDLINE: Intraocular Pressure, Glaucoma, Tonometry, Ocular. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

The purpose of this study was to compare iCare HOME (TA022) with Goldmann applanation tonometer and to evaluate the self-tonometry measurements among the Indian population.

Methods

Eligible patients underwent iCare HOME training through guided demonstration (verbal, pictorial, video) and practised self-tonometry measures using iCare HOME. Certification for independent iCare HOME measure was provided if first iCare HOME intraocular pressure (IOP) measurement fell within ± 5 mmHg of Goldmann applanation tonometer (GAT) measurement which was measured by the trained clinician (principal investigator). Certified participants underwent simulated home self-tonometry measurements using iCare HOME, and agreement with GAT IOP measurements was assessed.

Results

Seven of 83 participants (8.43%) failed to complete the study due to difficulty in performing the task, leading to non-certification. Patients who could use the iCare HOME had a mean age of 53 ± 15.55years (53% males; 46% females). Only one in 12 subjects did not qualify to use iCare HOME. The overall mean difference between iCare HOME and GAT was 0.83 mmHg (95%, 3.92 and -2.25) with 94.7% agreement. At various pressure ranges, 7-16 mmHg, 17-23 mmHg and >23 mmHg, the mean difference between iCare HOME and GAT was 1.22 mmHg (95%, 4.32 and -1.86) with 97.4% agreement, 0.77 mmHg (95%,3.69 and -2.19) with 91.3% agreement, -0.11 mmHg (95%, 2.52 and -2.74) with 100% agreement respectively. The intra-class correlation coefficient of the iCare HOME device was 0.997(95% CI,0.995-0.998).

Conclusions

Patients were able to perform self- tonometry using iCare HOME with good reliability and safety. iCare HOME shows good agreement with GAT. Further studies with proper home environment will be recommended to study the diurnal comparisons with GAT.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Intraocular PressureGlaucomaTonometry, OcularHumansMaleFemaleMiddle AgedAdultAgedSelf Care

Stichwörter der Autor:innen

GlaucomaGoldmann applanation tonometryIntraocular pressureiCare HOME

Autor:innen und Institutionen

  • Department of Allied Health Sciences,Faculty of Life & Allied Health Sciences(FLAHS), Ramaiah University of Applied Sciences, Bangalore, Karnataka, 560054, India.
  • Department of Optometry, Manipal College of Health Professions (MCHP), Manipal Academy of Higher Education (MAHE), Manipal, Udupi (Dist), Karnataka, 576104, India.
  • Department of Ophthalmology, Kasturba Medical College, Manipal, Manipal Academy of Higher Education (MAHE), Manipal, Udupi (Dist), Karnataka, 576104, India.

ORCID-Profile: S Ve Ramesh, Neetha I R Kuzhuppilly, Aditya Chaitanya

Zitieren

APA

Nayak, A., Ramesh, S. V., Kuzhuppilly, N. I. R., Pai, V. H., & Chaitanya, A. (2023). Performance of home-based self-tonometry (iCare HOME (TA022)) for measuring intraocular pressure among healthy and glaucoma patients. F1000Research, 12, 128. https://doi.org/10.12688/f1000research.123104.2

Vancouver

Nayak A, Ramesh SV, Kuzhuppilly NIR, Pai VH, Chaitanya A. Performance of home-based self-tonometry (iCare HOME (TA022)) for measuring intraocular pressure among healthy and glaucoma patients. F1000Research. 2023;12:128. doi:10.12688/f1000research.123104.2

BibTeX
@article{Nayak2023,
  author = {Nayak, Anush and Ramesh, S Ve and Kuzhuppilly, Neetha I R and Pai, Vijaya H and Chaitanya, Aditya},
  title = {Performance of home-based self-tonometry (iCare HOME (TA022)) for measuring intraocular pressure among healthy and glaucoma patients},
  journal = {F1000Research},
  year = {2023},
  volume = {12},
  pages = {128},
  doi = {10.12688/f1000research.123104.2}
}
Kostenlos starten →