Accuracy of a novel method for IOP measurement without applying pressure on the cornea
International Ophthalmology · 2024
DOI: 10.1007/s10792-024-03213-7 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Ophthalmology, 2024. Verfasst von Amit Biran und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Soraski Medical Center, Tel Aviv, Israel. Hauptthemen laut MEDLINE: Intraocular Pressure, Tonometry, Ocular, Cornea. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Technion - Israel Institute of Technology.
Abstract (Originalsprache)
Purpose
To assess the accuracy of a prototype novel instrument for intra ocular pressure (IOP) measurements not involving corneal pressure application.
Design
Prospective case control study.
Methods
An institutional study including 16 healthy volunteers without ocular pathology. IOP in both eyes of the participants was measured four times in different body positions with the novel prototype and reference instrument (Goldmann applanation tonometer (GAT) or iCare (iCare Finland OY, Vantaa, Finland)). IOP results were compared between the prototype and the reference instruments in 116 pairs of measurement.
Results
Overall no statistically significant difference was found between the presented prototype and the reference instrument. Stratifying measurements by instrument used revealed no significant difference for GAT and statistical significant (yet clinically insignificant) difference for iCare.
Conclusions
The presented prototype demonstrates good clinical agreement of IOP measuring results with reference instruments Further large-scale studies assessing this instrument in glaucoma patients are warranted.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Ophthalmology, Soraski Medical Center, Tel Aviv, Israel.
- Department of Applied Physics, Weizmann Institute of Science, Rehovot, Israel.
- Department of Ophthalmology, Rambam Health Care Campus Affiliated With the Bruce and Ruth Rappaport Faculty of Medicine, Technion-Israel Institute of Technology, POB 9602, 3109601, Haifa, Israel. michael@intername.co.il.
- Department of Ophthalmology, Rambam Health Care Campus Affiliated With the Bruce and Ruth Rappaport Faculty of Medicine, Technion-Israel Institute of Technology, POB 9602, 3109601, Haifa, Israel.
Förderung
- Technion - Israel Institute of Technology
Zitieren
APA
Biran, A., Sayar, M., Lazzar, M., Mimouni, M., & Beiran, I. (2024). Accuracy of a novel method for IOP measurement without applying pressure on the cornea. International Ophthalmology. https://doi.org/10.1007/s10792-024-03213-7
Vancouver
Biran A, Sayar M, Lazzar M, Mimouni M, Beiran I. Accuracy of a novel method for IOP measurement without applying pressure on the cornea. International Ophthalmology. 2024. doi:10.1007/s10792-024-03213-7
BibTeX
@article{Biran2024,
author = {Biran, Amit and Sayar, Mordehai and Lazzar, Moshe and Mimouni, Michael and Beiran, Itzchak},
title = {Accuracy of a novel method for IOP measurement without applying pressure on the cornea},
journal = {International Ophthalmology},
year = {2024},
doi = {10.1007/s10792-024-03213-7}
}
EUVIMED