Ablation apprentices and their first experience of pulmonary vein isolation procedure on paroxysmal atrial fibrillation with different sheaths
Journal of Cardiothoracic Surgery · 2024
DOI: 10.1186/s13019-024-02826-4 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Journal of Cardiothoracic Surgery, 2024. Verfasst von Ye Liu und 8 weiteren Autor:innen. Beteiligte Einrichtung: Heart Center & Beijing Key Laboratory of Hypertension, Beijing Chaoyang Hospital, Capital Medical University, Beijing, 100020, China. Hauptthemen laut MEDLINE: Atrial Fibrillation, Pulmonary Veins, Catheter Ablation. Die Arbeit nennt 12 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China.
Abstract (Originalsprache)
Objectives
This study aimed at exploring how using different kinds of sheaths will affect the very first ablation procedure of apprentices.
Methods
15 patients with paroxysmal atrial fibrillation were randomized to used fixed-curve, conventional steerable or visualized steerable sheath, and received complete isolation of pulmonary veins. All ablations were the very first procedure performed by 15 ablation apprentices. The use of fluoroscopy and catheter stability during the PVI were analyzed.
Results
Procedure duration was much longer in the fixed-curve group (116.8 ± 27 vs. 62.2 ± 17 vs. 60.4 ± 17, p < 0.001). X-ray exposure was lowest with visualized sheath (17.6 ± 5 vs. 18.6 ± 6 vs. 5.2 ± 6, p < 0.001). CF SD differed significantly, especially at the anterior aspect of LSPV (7.90 ± 2.90 vs. 5.04 ± 2.18 vs. 4.52 ± 2.40, p < 0.001) and posterior aspect of RSPV (6.84 ± 2.79 vs. 3.42 ± 2.04 vs. 3.50 ± 2.30, p < 0.001) in the fixed-curve group. Impedance drop was significantly smaller in the fixed-curve group at the anterior aspect of LSPV (8.74 ± 3.02 vs. 11.49 ± 5.48 vs. 12.57 ± 5.96, p = 0.005).
Conclusion
Even for the very first ablation procedure of an ablation apprentice, the use of steerable sheaths will significantly reduce the procedure duration and improve the catheter stability, but only visualized steerable sheath can reduce fluoroscopic time.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Heart Center & Beijing Key Laboratory of Hypertension, Beijing Chaoyang Hospital, Capital Medical University, Beijing, 100020, China.
- Heart Center & Beijing Key Laboratory of Hypertension, Beijing Chaoyang Hospital, Capital Medical University, Beijing, 100020, China. yangxinchun1957@yeah.net.
Förderung
- National Natural Science Foundation of China
Zitieren
APA
Liu, Y., Song, J., Wang, S., Liu, L., Liu, X., Liu, Z., Wang, Y., Zhao, L., & Yang, X. (2024). Ablation apprentices and their first experience of pulmonary vein isolation procedure on paroxysmal atrial fibrillation with different sheaths. Journal of Cardiothoracic Surgery. https://doi.org/10.1186/s13019-024-02826-4
Vancouver
Liu Y, Song J, Wang S, Liu L, Liu X, Liu Z, et al. Ablation apprentices and their first experience of pulmonary vein isolation procedure on paroxysmal atrial fibrillation with different sheaths. Journal of Cardiothoracic Surgery. 2024. doi:10.1186/s13019-024-02826-4
BibTeX
@article{Liu2024,
author = {Liu, Ye and Song, Jingjing and Wang, Siyu and Liu, Lifeng and Liu, Xiaoqing and Liu, Zheng and Wang, Yuxing and Zhao, Lei and Yang, Xinchun},
title = {Ablation apprentices and their first experience of pulmonary vein isolation procedure on paroxysmal atrial fibrillation with different sheaths},
journal = {Journal of Cardiothoracic Surgery},
year = {2024},
doi = {10.1186/s13019-024-02826-4}
}
EUVIMED