Procedural efficiency is enhanced combining the pentaspline pulsed field ablation catheter with three-dimensional electroanatomical mapping system for pulmonary vein isolation
Journal of Interventional Cardiac Electrophysiology · 10.06.2024 · 67 (9) · 1993-2001
DOI: 10.1007/s10840-024-01846-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Interventional Cardiac Electrophysiology, online erschienen am 10.06.2024. Verfasst von Alessio Falasca Zamponi und 5 weiteren Autor:innen. Beteiligte Einrichtung: Capio Arytmicenter Stockholm AB, Stockholm, Sweden. alessio.fz@tuta.io. Hauptthemen laut MEDLINE: Atrial Fibrillation, Pulmonary Veins, Catheter Ablation, Imaging, Three-Dimensional. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Karolinska Institute.
Abstract (Originalsprache)
Background
Pulsed field ablation (PFA) offers a safe, non-thermal alternative for pulmonary vein isolation (PVI) in patients with atrial fibrillation (AF). Currently, the pentaspline PFA-system relies heavily on fluoroscopy for catheter manipulation, which poses challenges due to the complexity of left atrium anatomy. Incorporating three-dimensional electroanatomical mapping (3D-EAM) could improve procedural efficiency reducing dependency on fluoroscopy guidance. This study aims to evaluate the effects of integration of 3D-EAM with PFA during PVI.
Methods
Between September 2022 and December 2023, we retrospectively enrolled 248 patients with paroxysmal or persistent AF undergoing PVI at our center using the pentaspline PFA catheter. The control group ( n = 104) received conventional PFA with fluoroscopic guidance alone, while the intervention group ( n = 144) underwent PVI with PFA with 3D-EAM integration. Primary outcomes were procedural time, fluoroscopy time (FT), and dose area product (DAP). Secondary endpoints included acute procedural success and incidence of periprocedural complications.
Results
In the 3D-EAM-PFA group, procedural time was 63.3 ± 14.3 min, compared to 65.6 ± 14.9 min in the control group ( p = 0.22). The 3D-EAM group experienced significantly reduced FT (9.7 ± 4.4 min vs. 16.7 ± 5.2 min) and DAP (119.2 ± 121.7 cGycm 2 vs. 338.7 ± 229.9 cGycm 2 ) compared to the control group, respectively ( p < 0.001). Acute procedural success was achieved in all cases. No major complications were observed in either group.
Conclusion
Integration of 3D-EAM with the pentaspline PFA catheter for PVI in AF treatment offers a promising approach, with significantly reduced fluoroscopy exposure without compromising procedural time and efficacy. Graphical abstract
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Capio Arytmicenter Stockholm AB, Stockholm, Sweden. alessio.fz@tuta.io.
- Department of Clinical Science and Education, Division of Cardiology, Karolinska Institutet, South Hospital, Stockholm, Sweden. alessio.fz@tuta.io.
- Capio Arytmicenter Stockholm AB, Stockholm, Sweden.
- Department of Clinical Science and Education, Division of Cardiology, Karolinska Institutet, South Hospital, Stockholm, Sweden.
Förderung
- Karolinska Institute
Zitieren
APA
Falasca Zamponi, A., Olson, J., Scheel, S., Englund, A., Scorza, R., & Tabrizi, F. (2024). Procedural efficiency is enhanced combining the pentaspline pulsed field ablation catheter with three-dimensional electroanatomical mapping system for pulmonary vein isolation. Journal of Interventional Cardiac Electrophysiology, 67(9), 1993-2001. https://doi.org/10.1007/s10840-024-01846-3
Vancouver
Falasca Zamponi A, Olson J, Scheel S, Englund A, Scorza R, Tabrizi F. Procedural efficiency is enhanced combining the pentaspline pulsed field ablation catheter with three-dimensional electroanatomical mapping system for pulmonary vein isolation. Journal of Interventional Cardiac Electrophysiology. 2024;67(9):1993-2001. doi:10.1007/s10840-024-01846-3
BibTeX
@article{FalascaZamponi2024,
author = {Falasca Zamponi, Alessio and Olson, Jens and Scheel, Sergej and Englund, Anders and Scorza, Raffaele and Tabrizi, Fariborz},
title = {Procedural efficiency is enhanced combining the pentaspline pulsed field ablation catheter with three-dimensional electroanatomical mapping system for pulmonary vein isolation},
journal = {Journal of Interventional Cardiac Electrophysiology},
year = {2024},
volume = {67},
number = {9},
pages = {1993--2001},
doi = {10.1007/s10840-024-01846-3}
}
EUVIMED