Kurz gefasst

Fachartikel in der Zeitschrift The Neuroradiology Journal, 2023. Verfasst von Emanuele Camerucci und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Mayo Clinic, Rochester, MN, USA. Hauptthemen laut MEDLINE: Artifacts, Hydrocephalus, Normal Pressure. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 6 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and purpose : Post-shunt MRI is usually performed at 1.5T under the general assumption that shunt-related susceptibility artifacts would be greater at higher field strengths.

Purpose

The purpose is to show that imaging post-shunt idiopathic normal pressure hydrocephalus (iNPH) patients at 3T is feasible and with reduced artifacts as compared to 1.5T.

Methods

We manually measured transverse dimensions of artifact at the levels of lateral ventricles, cerebral aqueduct, and cerebellar hemisphere. Areas/volumes of artifacts were calculated assuming an elliptic/ellipsoid shape. Relative extent of shunt-related artifact between field strengths was rated by 3 readers on a 5-point Likert scale. A Wilcoxon Signed Rank Test was used to compare artifact at 1.5T vs 3T for each sequence, with a significance level set at p < 0.05.

Results

Artifact areas were calculated in 22 iNPH patients; artifacts were on average smaller at 3T vs 1.5T on MPRAGE, DWI, and GRE sequences. On T2 FLAIR and T2 FSE, artifacts at 3T were larger than 1.5T. On the qualitative analysis, artifact effects were less at 3T vs 1.5T on DWI, greater at 3T on T2 FSE, and had mixed results on GRE.

Conclusion

Our results indicate feasibility of post-shunt imaging with the CERTAS Plus valve at 3T based on shunt-related artifact that is less than or equal in extent to that on 1.5T on most standard clinical imaging sequences. Our findings, corroborated by the qualitative image review, suggest that dedicated clinical imaging sequences for devices may allow for reduction in artifact extent at both 1.5T and 3T.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ArtifactsHydrocephalus, Normal PressureHumansMagnetic Resonance ImagingImage Interpretation, Computer-Assisted

Stichwörter der Autor:innen

CERTAS Plus valveNPHsusceptibility artifact

Autor:innen und Institutionen

  • Department of Radiology, Mayo Clinic, Rochester, MN, USA.
  • Department of Neurologic Surgery, Mayo Clinic, Rochester, MN, USA.
  • Department of Physiology and Biomedical Engineering, Mayo Clinic, Rochester, MN, USA.
  • Department of Neurology, Mayo Clinic, Rochester, MN, USA.

ORCID-Profile: Emanuele Camerucci, Jeremy K Cutsforth-Gregory

Zitieren

APA

Camerucci, E., Elder, B. D., Shu, Y., Messina, S. A., Gunter, J. L., Graff-Radford, J., Jones, D. T., Botha, H., Cutsforth-Gregory, J. K., Jack, C. R., Huston, J., & Cogswell, P. M. (2023). Field strength difference in extent of artifacts induced by CERTAS Plus valves in patients with idiopathic normal pressure hydrocephalus. The Neuroradiology Journal. https://doi.org/10.1177/19714009231173099

Vancouver

Camerucci E, Elder BD, Shu Y, Messina SA, Gunter JL, Graff-Radford J, et al. Field strength difference in extent of artifacts induced by CERTAS Plus valves in patients with idiopathic normal pressure hydrocephalus. The Neuroradiology Journal. 2023. doi:10.1177/19714009231173099

BibTeX
@article{Camerucci2023,
  author = {Camerucci, Emanuele and Elder, Benjamin D and Shu, Yunhong and Messina, Steven A and Gunter, Jeffrey L and Graff-Radford, Jonathan and Jones, David T and Botha, Hugo and Cutsforth-Gregory, Jeremy K and Jack, Clifford R and Huston, John and Cogswell, Petrice M},
  title = {Field strength difference in extent of artifacts induced by CERTAS Plus valves in patients with idiopathic normal pressure hydrocephalus},
  journal = {The Neuroradiology Journal},
  year = {2023},
  doi = {10.1177/19714009231173099}
}
Kostenlos starten →