Differences in Arterial Oxygen Saturation Measurements Between Two Co-Oximeters From Different Manufacturers

Tyler J Law iD, Gregory Leeb, Xander M R van Wijk, Robert J Kopotic, Carolyn M Hendrickson, Ella Behnke, Philip E Bickler, John Feiner, René Vargas Zamora, Michael S Lipnick

Respiratory Care · 2025

DOI: 10.1177/19433654251376285 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Respiratory Care, 2025. Verfasst von Tyler J Law und 9 weiteren Autor:innen. Beteiligte Einrichtung: Drs. Law, Leeb, Bickler, Feiner, Lipnick, and Mr. Vargas Zamora are affiliated with Department of Anesthesia & Perioperative Care, University of Cali…. Hauptthemen laut MEDLINE: Oximetry, Oxygen Saturation, Oxygen. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Functional arterial hemoglobin oxygen saturation is a critical clinical parameter and the reference for verifying accuracy of S pO 2 values during performance testing of a pulse oximeter. Variable performance between blood gas analyzers that perform co-oximetry may complicate interpretation of research studies and regulatory guidance. This study quantifies systematic differences between multiple same model analyzers from two widely used brands of co-oximeters.

Methods

Three sequential experiments were conducted using controlled hypoxemia in healthy adult volunteers according to regulatory clearance criteria for use of functional arterial oxygenation from co-oximetry analysis (sO 2 ), to simultaneous S pO 2 values over the range of 70–100%. To account for potential methodological confounders in experiments 1 and 2, a third experiment analyzed 31 arterial samples across multiple same model analyzers for co-oximetry (two Werfen GEM Premier 5000, and three Radiometer ABL90 Flex) in a randomized order after simultaneous preparation. Statistical analysis included paired differences, mixed-effects modeling, and variance analysis to isolate analyzer-specific effects while controlling for confounding variables.

Results

The GEM analyzers consistently measured sO 2 2.3% higher on average than ABL analyzers (95% CI 2.01–2.63%, P < .001). This difference was more pronounced at lower saturations (3.2% in the sO 2 70s range vs 1.4% in the 90s range). Despite strong correlation between measurements (r = 0.999), 96.8% of paired samples differed by >1% and 61.3% by >2%. Within-manufacturer variability was similar between brands (mean absolute difference: ABL 0.38%, GEM 0.47%, P = .55).

Conclusions

Clinically important systematic differences in sO 2 measurement exist between GEM and ABL blood analyzers that perform co-oximetry, with potential implications for patient care and pulse oximeter verification studies. Larger studies with more co-oximeter brands, coupled with blood tonometry, are needed to better characterize these findings. In the meantime, regulatory standards and reports of pulse oximeter clinical trials should account for these findings and specify which manufacturer’s co-oximeters were used.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

OximetryOxygen SaturationOxygenHumansAdultMaleFemaleHypoxiaReproducibility of ResultsHealthy Volunteers

Stichwörter der Autor:innen

blood gas analysiscomparative studyhealthy volunteershypoxemiamonitoring (physiologic)oximetryreproducibility of resultstechnology assessment (biomedical)

Substanzen

Oxygen

Autor:innen und Institutionen

  • Drs. Law, Leeb, Bickler, Feiner, Lipnick, and Mr. Vargas Zamora are affiliated with Department of Anesthesia & Perioperative Care, University of California, San Francisco, California, USA.
  • Drs. Law, Leeb, Lipnick, and Mr. Vargas Zamora are affiliated with Center for Health Equity in Surgery and Anesthesia, University of California, San Francisco, California, USA.
  • Drs. Law, Leeb, Lipnick, Bickler, Feiner, Ms. Behnke, and Mr. Vargas Zamora are affiliated with Hypoxia Research Laboratory, University of California, San Francisco, California, USA.
  • Dr. van Wijk is affiliated with Department of Laboratory Medicine, University of California, San Francisco, California, USA.
  • Dr. Kopotic is affiliated with the Joint Working Group as the Oximeters Convener for medical device standards.
  • Dr. Hendrickson is affiliated with Division of Pulmonary, Critical Care, Allergy and Sleep Medicine, University of California, San Francisco, California, USA.

ORCID-Profile: Tyler J Law

Zitieren

APA

Law, T. J., Leeb, G., van Wijk, X. M. R., Kopotic, R. J., Hendrickson, C. M., Behnke, E., Bickler, P. E., Feiner, J., Vargas Zamora, R., & Lipnick, M. S. (2025). Differences in Arterial Oxygen Saturation Measurements Between Two Co-Oximeters From Different Manufacturers. Respiratory Care. https://doi.org/10.1177/19433654251376285

Vancouver

Law TJ, Leeb G, van Wijk XMR, Kopotic RJ, Hendrickson CM, Behnke E, et al. Differences in Arterial Oxygen Saturation Measurements Between Two Co-Oximeters From Different Manufacturers. Respiratory Care. 2025. doi:10.1177/19433654251376285

BibTeX
@article{Law2025,
  author = {Law, Tyler J and Leeb, Gregory and van Wijk, Xander M R and Kopotic, Robert J and Hendrickson, Carolyn M and Behnke, Ella and Bickler, Philip E and Feiner, John and Vargas Zamora, René and Lipnick, Michael S},
  title = {Differences in Arterial Oxygen Saturation Measurements Between Two Co-Oximeters From Different Manufacturers},
  journal = {Respiratory Care},
  year = {2025},
  doi = {10.1177/19433654251376285}
}
Kostenlos starten →