Continuous Glucose Monitoring Accuracy With In Vivo Exposure to Magnetic Resonance Imaging

Ray Wang iD, Wen Phei Choong, Shana Woodthorpe, Mervyn Kyi, Spiros Fourlanos

Journal of Diabetes Science and Technology · 2025-01 · 19 (1) · 265-266

DOI: 10.1177/19322968241289446 ↗

Kurz gefasst

Leserbrief in der Zeitschrift Journal of Diabetes Science and Technology, online erschienen am 16.10.2024. Verfasst von Ray Wang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diabetes and Endocrinology, The Royal Melbourne Hospital, Parkville, VIC, Australia. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 5 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Australian Commonwealth Government, University of Melbourne.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Removal of wearable transcutaneous continuous glucose monitoring (CGM) devices is recommended before magnetic resonance imaging (MRI), due to potential safety and accuracy concerns.We report two cases of in vivo CGM use during MRI, and describe subsequent CGM accuracy.A 74-year-old woman (case 1) with type 1 diabetes (T1D) inadvertently wore her Dexcom G6 CGM (DexCom Inc, San Diego, CA, USA), located on her abdomen, during a 25-minute cervical spine MRI (Magnetom Aera 1.5T; Siemens, Munich, Germany).Another 54-year-old woman (case 2) with T1D inadvertently wore her FreeStyle Libre 2 CGM (Abbott Diabetes Care, Doncaster, VIC, Australia), located on her arm, during a 15-minute lumbar spine MRI (Magnetom Sola 1.5T; Siemens, Munich, Germany) and subsequent 26-minute brain MRI (Magnetom Vida 3T; Siemens, Munich, Germany).In both cases, CGM devices did not dislodge during MRI, and continued to record continuous glucose readings during, and following, MRI (Figure 1a).To assess CGM accuracy, we time-matched (±7 minutes) CGM glucose with point-of-care (POC) capillary blood glucose (BG) measures (Nova StatStrip; Nova Biomedical, Waltham, MA, USA) from the 72 hours before and after MRI, after excluding measures outside device reportable ranges.Twenty-five time-matched BG pairs (21 pre-MRI; four post-MRI) from case 1 (mean BG 229 mg/dL, SD 65) resulted in mean absolute relative difference (MARD) of 6.7% pre-MRI and 7.9% post-MRI.For case 2, six post-MRI BG pairs (four after first MRI; two after second MRI), but no pre-MRI BG pairs, were obtained (mean BG 282 mg/dL, SD 82), with MARD 5.0% post-MRI.All BG pairs were within Consensus Error Grid zones A/B (93.5% zone A; 6.5% zone B) (Figure 1b). 1 Currently, the only CGM labeled MRI-conditional is the implantable subcutaneous Eversense sensor (Ascensia Diabetes Care, Parsippany, NJ, USA). 2 Wearable transcutaneous CGM 1289446D STXXX10.

Schlagwörter

Stichwörter der Autor:innen

CGMCGMSMRImagnetic resonance studiessensor accuracytype 1 diabetes

Autor:innen und Institutionen

  • Department of Diabetes and Endocrinology, The Royal Melbourne Hospital, Parkville, VIC, Australia.
  • Department of Medicine, Melbourne Medical School, The University of Melbourne, Parkville, VIC, Australia.
  • Australian Centre for Accelerating Diabetes Innovations, Melbourne Medical School, The University of Melbourne, Parkville, VIC, Australia.

ORCID-Profile: Ray Wang

Förderung

  • Australian Commonwealth Government
  • University of Melbourne

Zitieren

APA

Wang, R., Choong, W. P., Woodthorpe, S., Kyi, M., & Fourlanos, S. (2025). Continuous Glucose Monitoring Accuracy With In Vivo Exposure to Magnetic Resonance Imaging. Journal of Diabetes Science and Technology, 19(1), 265-266. https://doi.org/10.1177/19322968241289446

Vancouver

Wang R, Choong WP, Woodthorpe S, Kyi M, Fourlanos S. Continuous Glucose Monitoring Accuracy With In Vivo Exposure to Magnetic Resonance Imaging. Journal of Diabetes Science and Technology. 2025;19(1):265-266. doi:10.1177/19322968241289446

BibTeX
@article{Wang2025,
  author = {Wang, Ray and Choong, Wen Phei and Woodthorpe, Shana and Kyi, Mervyn and Fourlanos, Spiros},
  title = {Continuous Glucose Monitoring Accuracy With In Vivo Exposure to Magnetic Resonance Imaging},
  journal = {Journal of Diabetes Science and Technology},
  year = {2025},
  volume = {19},
  number = {1},
  pages = {265--266},
  doi = {10.1177/19322968241289446}
}
Kostenlos starten →