Real‐World Accuracy of a Continuous Glucose Monitoring System after Radiologic Exposure
Pediatric Diabetes · 2024-01 · 2024 (1)
DOI: 10.1155/2024/2210509 ↗Kurz gefasst
Fachartikel in der Zeitschrift Pediatric Diabetes, online erschienen am 07.08.2024. Verfasst von Siobhan Tellez und 8 weiteren Autor:innen. Beteiligte Einrichtung: Division of Endocrinology Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio, USA. Hauptthemen laut MEDLINE: Blood Glucose Self-Monitoring, Blood Glucose, Diabetes Mellitus, Type 1, Radiation Exposure. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Dexcom.
Abstract (Originalsprache)
Background
The increasing use of continuous glucose monitor (CGM) necessitates a review of variables that impact accuracy and interrupt use. Manufacturer recommendations include removing CGMs before diagnostic imaging, such as X‐ray and computed tomography (CT). Early removal and replacement of CGM components present financial, clinical, and psychosocial burdens to the wearer and interrupt optimal management of diabetes for pediatric patients who receive a total pancreatectomy with islet autotransplantation (TPIAT). The study’s aim was to evaluate the effect of scatter dose exposure during X‐ray or CT if the CGM remained intact but outside the field of view (FoV). Materials and Methods . Participants were followed through the first 3 months after TPIAT surgery, managed diabetes with an insulin pump and CGM, and were routinely exposed to diagnostic imaging. Participants’ CGMs were unshielded by a protective apron during any X‐ray or CT procedures for the duration of the study period, and the transmitter was collected after expiration or removal. Glucometer data was collected from hospital records and home glucometer downloads. Mixed models were used to analyze absolute differences between matched CGM and glucometer values, and Clarke error grid analyses (EGA) were performed. Scatter dose exposure was derived using anthropomorphic phantoms and calculated retrospectively.
Results
A total of 14 patients (median 12.2 years, 64% female) received a median of five diagnostic imaging procedures with a median cumulative scatter dose of 559 µ Gy. The absolute difference between the CGM and glucometer values was not significantly associated with the cumulative scatter dose ( p = 0.17) or time from TPIAT ( p = 0.24) when analyzed in a mixed model. Regardless of scatter dose exposure, time from TPIAT, or glucometer, ≥98% of glucose values fell within zones A and B on EGA.
Conclusion
Scatter dose exposure from diagnostic imaging did not affect the clinical accuracy of CGM values for the duration of transmitter use. Leaving CGM components in place when not in the FoV during diagnostic imaging successfully mitigated interruptions to use and undue burden or cost to participants.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Division of Endocrinology Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio, USA.
- Division of Biostatistics and Epidemiology Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio, USA.
- Department of Radiology Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio, USA.
- Department of Radiology University of Cincinnati College of Medicine, Cincinnati, Ohio, USA.
- Department of Pediatrics University of Cincinnati College of Medicine, Cincinnati, Ohio, USA.
- Division of Pediatric General and Thoracic Surgery Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio, USA.
ORCID-Profile: Siobhan Tellez, Lindsey Hornung, Emily Smith, Andrew Trout, Samuel Brady, Colleen Lowe, Deborah Elder
Förderung
- Dexcom
Zitieren
APA
Tellez, S., Hornung, L., Smith, E., Trout, A., Brady, S., Lowe, C., Courter, J., Abu-El-Haija, M., & Elder, D. (2024). Real‐World Accuracy of a Continuous Glucose Monitoring System after Radiologic Exposure. Pediatric Diabetes, 2024(1). https://doi.org/10.1155/2024/2210509
Vancouver
Tellez S, Hornung L, Smith E, Trout A, Brady S, Lowe C, et al. Real‐World Accuracy of a Continuous Glucose Monitoring System after Radiologic Exposure. Pediatric Diabetes. 2024;2024(1). doi:10.1155/2024/2210509
BibTeX
@article{Tellez2024,
author = {Tellez, Siobhan and Hornung, Lindsey and Smith, Emily and Trout, Andrew and Brady, Samuel and Lowe, Colleen and Courter, Joshua and Abu-El-Haija, Maisam and Elder, Deborah},
title = {Real‐World Accuracy of a Continuous Glucose Monitoring System after Radiologic Exposure},
journal = {Pediatric Diabetes},
year = {2024},
volume = {2024},
number = {1},
doi = {10.1155/2024/2210509}
}
EUVIMED