Impact of X-Ray Exposure From Computed Tomography on Wearable Insulin Delivery Devices
Journal of Diabetes Science and Technology · 2024-11 · 18 (6) · 1387-1391
DOI: 10.1177/19322968231169722 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Diabetes Science and Technology, online erschienen am 26.04.2023. Verfasst von Frank Dong und 5 weiteren Autor:innen. Beteiligte Einrichtung: The University of Texas MD Anderson Cancer Center, Houston, TX, USA. Hauptthemen laut MEDLINE: Insulin Infusion Systems, Tomography, X-Ray Computed, Wearable Electronic Devices, Insulin. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 9 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Insulet Corporation.
Abstract (Originalsprache)
Background
To investigate the impact of radiation exposure from a computed tomography (CT) scanner on the functional integrity of a wearable insulin delivery system.
Methods
A total of 160 Omnipods and four personal diabetes managers (PDMs) were evenly divided into four groups: (1) control group (no radiation exposure), (2) typical radiation exposure group, (3) 4× typical radiation exposure group, and (4) scatter radiation group. Pods were attached to an anthropomorphic torso phantom on the abdomen (direct irradiation) or shoulder (scatter radiation) region. A third-generation dual-source CT scanner was used to scan the pods using either a typical exposure (used for routine CT abdominal study of a median size patient) or 4× typical exposure. A manufacturer-recommended 20-step functionality test was performed for all 160 Omnipods.
Results
The radiation dose (measured in volume CT Dose index) was 16 mGy for a typical exposure, and 64 mGy for 4× typical exposure. The scatter radiation is less than 0.1 mGy. All Pods passed the functionality test except one pod in the scatter radiation group, which sounded an alarm due to occlusion. The blockage to the fluid was due to a kink in the soft cannula, a mechanical issue not caused by the radiation exposure.
Conclusions
This study suggests X-ray exposure levels used in radiological imaging procedures do not negatively impact the functional integrity of Omnipods. This finding may support the potential for the manufacturer to remove the warning that patients should remove the Pod for X-ray imaging procedures, which will have a huge impact on patient care.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- The University of Texas MD Anderson Cancer Center, Houston, TX, USA.
- Imaging Institute, Cleveland Clinic, Cleveland, OH, USA.
- Insulet Corporation, Acton, MA, USA.
ORCID-Profile: Frank Dong
Förderung
- Insulet Corporation
Zitieren
APA
Dong, F., Johnson, P., Fong, G., Nguyen, A., Lauand, F., & Vienneau, T. (2024). Impact of X-Ray Exposure From Computed Tomography on Wearable Insulin Delivery Devices. Journal of Diabetes Science and Technology, 18(6), 1387-1391. https://doi.org/10.1177/19322968231169722
Vancouver
Dong F, Johnson P, Fong G, Nguyen A, Lauand F, Vienneau T. Impact of X-Ray Exposure From Computed Tomography on Wearable Insulin Delivery Devices. Journal of Diabetes Science and Technology. 2024;18(6):1387-1391. doi:10.1177/19322968231169722
BibTeX
@article{Dong2024,
author = {Dong, Frank and Johnson, Paul and Fong, Grant and Nguyen, Alex and Lauand, Felipe and Vienneau, Todd},
title = {Impact of X-Ray Exposure From Computed Tomography on Wearable Insulin Delivery Devices},
journal = {Journal of Diabetes Science and Technology},
year = {2024},
volume = {18},
number = {6},
pages = {1387--1391},
doi = {10.1177/19322968231169722}
}
EUVIMED