Evaluation of overexposure risk when there is a space between the subject and the couch in computed tomography: a phantom study
Radiological Physics and Technology · 2024
DOI: 10.1007/s12194-024-00804-y ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiological Physics and Technology, 2024. Verfasst von Yuta Fujiwara und 7 weiteren Autor:innen. Beteiligte Einrichtung: Division of Clinical Radiology Service, Okayama Central Hospital, 6-3, Ishimakitamachi, Kitaku, Okayama, 700-0017, Japan. y_fujiwara_gishi@yahoo.co.j…. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Phantoms, Imaging. Die Arbeit nennt 13 Literaturangaben.
Abstract (Originalsprache)
The purpose of this study was to investigate the risk of overexposure associated with automatic tube current modulation (ATCM) and automatic couch height positioning compensation mechanism (AHC) in computed tomography (CT) systems, particularly in scenarios involving a gap between the subject and the couch. Results revealed that when AHC was enabled, CT dose index volume (CTDIvol) increased by approximately 10% at 2.5 cm, 20% at 5.0 cm, and 40% at 10.0 cm gaps compared to close contact conditions. While the AHC function ensures consistent exposure doses and image quality regardless of subject positioning relative to the CT gantry isocenter, the study highlights a potential risk of overexposure when a gap exists between the subject and the couch. These findings offer valuable insights for optimizing CT imaging protocols and underscore the importance of carefully considering subject positioning in clinical practice.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Clinical Radiology Service, Okayama Central Hospital, 6-3, Ishimakitamachi, Kitaku, Okayama, 700-0017, Japan. y_fujiwara_gishi@yahoo.co.jp.
- Division of Clinical Radiology Service, Okayama Central Hospital, 6-3, Ishimakitamachi, Kitaku, Okayama, 700-0017, Japan.
- Department of Radiological Technology, Faculty of Health Science and Technology, Kawasaki University of Medical Welfare, Matsushima, Kurashiki, Okayama, 288701-0193, Japan.
Zitieren
APA
Fujiwara, Y., Kamihoriuchi, Y., Higuchi, F., Nakayama, S., Ohyama, Y., Sasaki, T., Watanabe, S., & Masuda, T. (2024). Evaluation of overexposure risk when there is a space between the subject and the couch in computed tomography: a phantom study. Radiological Physics and Technology. https://doi.org/10.1007/s12194-024-00804-y
Vancouver
Fujiwara Y, Kamihoriuchi Y, Higuchi F, Nakayama S, Ohyama Y, Sasaki T, et al. Evaluation of overexposure risk when there is a space between the subject and the couch in computed tomography: a phantom study. Radiological Physics and Technology. 2024. doi:10.1007/s12194-024-00804-y
BibTeX
@article{Fujiwara2024,
author = {Fujiwara, Yuta and Kamihoriuchi, Yoshiki and Higuchi, Fumie and Nakayama, Shinichi and Ohyama, Yutako and Sasaki, Tomoko and Watanabe, Shinsaku and Masuda, Takanori},
title = {Evaluation of overexposure risk when there is a space between the subject and the couch in computed tomography: a phantom study},
journal = {Radiological Physics and Technology},
year = {2024},
doi = {10.1007/s12194-024-00804-y}
}
EUVIMED