Loose Screw or No Screw? A Biomechanical Study to Answer a Common Intraoperative Dilemma
HAND · 30.10.2026
DOI: 10.1177/15589447261496536 ↗Kurz gefasst
Fachartikel in der Zeitschrift HAND, online erschienen am 30.10.2026. Verfasst von Christopher C Y Chung und 5 weiteren Autor:innen. Beteiligte Einrichtung: Stanford University, CA, USA. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Acumed, Inc.
Abstract (Originalsprache)
Background
Insertion of cortical screws with poor purchase during metacarpal plate fixation may occur, yet the biomechanical tradeoff of retaining versus removing a loose screw remains difficult to quantify. Removing a loose screw avoids risking screw migration but may decrease fixation strength or create a stress riser at the empty hole. Our study evaluated bending and torsional properties in plate fixation for comminuted metacarpal diaphyseal fractures with or without a single loose screw.
Methods
Sixty-four cadaveric fractures were fixed in bridge mode using nonlocking screws. A single screw was overtightened and stripped to simulate screw loosening and then either retained or removed. Testing groups were assigned based on loading configuration (cantilever bending or torsion) and the presence of a loose screw. Multivariable linear regression models evaluated relationships between screw presence and biomechanical outcome measures. Failure modes were analyzed using Fisher exact test.
Results
Cantilever bending load at 2 mm displacement was 7.6 N (SD = 1.9 N) with loose screw retained versus 8.0 N (SD = 2.6 N) with screw removed ( P = .796). Torque at 10° rotational displacement showed no difference between retained (0.363 N m, SD = 0.054 N m) and removed screw groups (0.370 N m, SD = 0.139 N m; P = .858). No differences were identified in stress or stiffness. Cantilever bending clinical failure occurred with all specimens bending through the plate. During torsional testing, 6 of 16 with retained loose screw secondarily failed, with the loose screw hole fracturing after rotating past the clinical endpoint.
Conclusion
Removing a single loose screw in bridge plate fixation for comminuted metacarpal diaphyseal fractures safely avoids screw migration without imparting a stress riser or compromising biomechanical strength.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Stanford University, CA, USA.
ORCID-Profile: Christopher C Y Chung, Nicole S Pham
Förderung
- Acumed, Inc
Zitieren
APA
Chung, C. C. Y., Harri, A., Shehadeh, K., Chan, C. K., Pham, N. S., & Yao, J. (2026). Loose Screw or No Screw? A Biomechanical Study to Answer a Common Intraoperative Dilemma. HAND. https://doi.org/10.1177/15589447261496536
Vancouver
Chung CCY, Harri A, Shehadeh K, Chan CK, Pham NS, Yao J. Loose Screw or No Screw? A Biomechanical Study to Answer a Common Intraoperative Dilemma. HAND. 2026. doi:10.1177/15589447261496536
BibTeX
@article{Chung2026,
author = {Chung, Christopher C Y and Harri, Adina and Shehadeh, Katherine and Chan, Calvin K and Pham, Nicole S and Yao, Jeffrey},
title = {Loose Screw or No Screw? A Biomechanical Study to Answer a Common Intraoperative Dilemma},
journal = {HAND},
year = {2026},
doi = {10.1177/15589447261496536}
}
EUVIMED