EUVIMEDEUVIMED entdecken
Crossref

Length of the fractured level screw has no significant effect on biomechanical stability in short-segment thoracolumbar fixation—in vitro biomechanical study in a sheep model

Selahaddin Aydemir, Anıl Murat Öztürk, Ahmet Karakaşlı, Reşit Buğra Hüsemoğlu, Mustafa Çeltik, Onur Süer, Ömer Akçalı

Acta Orthopaedica et Traumatologica Turcica · 23.10.2026 · 60 (5)

DOI: 10.5152/j.aott.2026.261021 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Acta Orthopaedica et Traumatologica Turcica, online erschienen am 23.10.2026. Verfasst von Selahaddin Aydemir und 6 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

While pedicle screw placement at the fracture level improves stability in short-segment posterior instrumentation for thoracolumbar fractures, the optimal screw length remains controversial. The biomechanical effect of screw length at the fracture level has not been clearly defined under controlled conditions. This study compared short (≤50% vertebral body length) and long (≥70%) screws in a sheep burst fracture model.

Methods

Twenty-one fresh sheep thoracolumbar spines with experimentally induced L1 burst fractures (AO Type A4) were randomly divided into 3 groups: Group 1 (no fracture-level screw), Group 2 (short screws: ≤ 50% of the vertebral body length [20 mm]), and Group 3 (long screws: ≥ 70% of the body length [35 mm]). All specimens were instrumented with pedicle screw constructs. Stability was assessed under axial compression, flexion–extension, lateral bending, and axial rotation using a Shimadzu AG-X testing machine (Kyoto, Japan). Axial displacement was recorded to derive range of motion (ROM) as a primary indicator of construct stability. Post-hoc power analysis revealed 58% power to detect a 0.5 mm displacement difference; therefore, results are interpreted as “no statistically significant difference” rather than equivalence.

Results

Incorporating a fracture-level screw significantly improved stability vs. Group 1 (P < .01). No significant differences were observed between short and long screws in axial compression ROM (0.91 ± 0.19 mm vs. 0.89 ± 0.26 mm, P = .87), flexion/extension ROM (1.41 ± 0.20 mm [3.2° ± 0.5°] vs. 1.17 ± 0.24 mm [2.7° ± 0.6°], P = .12), lateral bending ROM (1.27 ± 0.31 mm [2.9° ± 0.7°] vs. 0.88 ± 0.11 mm [2.0° ± 0.3°], P = .09), or axial rotation failure displacement (80.44 ± 41.31 mm vs. 71.44 ± 29.51 mm, P = .23). Fragment displacement was significantly greater in the long-screw group (3.51 ± 1.25 mm) than in the short-screw group (0.00 ± 0.00 mm, P < .001).

Conclusion

Within the limitations of this sheep model and with only 58% power to detect a 0.5 mm difference, short screws (≤50% vertebral body length) demonstrated no statistically significant difference in biomechanical stability compared to long screws (≥70%) in short segment posterior fixation under the tested conditions. However, long screws were associated with significant anterior fragment displacement, a potentially clinically relevant finding requiring further computed tomography (CT)-based validation. These findings suggest a potential advantage of short screws in avoiding anterior fragment displacement without compromising biomechanical stability. However, given the study’s power limitations and the absence of CT-based fragment displacement validation, these results are preliminary. Clinical recommendations require confirmation in adequately powered human studies with formal equivalence testing and post-instrumentation CT imaging. Cite this article as: Aydemir S, Ozturk AM, Karakasli A, et al. Length of the fractured level screw has no significant effect on biomechanical stability in shortsegment thoracolumbar fixation—in vitro biomechanical study in a sheep model. Acta Orthop Traumatol Turc., 2026; 60(5), 01021, doi: 10.5152/j.aott.2026.261021.

Zitieren

APA

Aydemir, S., Öztürk, A. M., Karakaşlı, A., Hüsemoğlu, R. B., Çeltik, M., Süer, O., & Akçalı, Ö. (2026). Length of the fractured level screw has no significant effect on biomechanical stability in short-segment thoracolumbar fixation—in vitro biomechanical study in a sheep model. Acta Orthopaedica et Traumatologica Turcica, 60(5). https://doi.org/10.5152/j.aott.2026.261021

Vancouver

Aydemir S, Öztürk AM, Karakaşlı A, Hüsemoğlu RB, Çeltik M, Süer O, et al. Length of the fractured level screw has no significant effect on biomechanical stability in short-segment thoracolumbar fixation—in vitro biomechanical study in a sheep model. Acta Orthopaedica et Traumatologica Turcica. 2026;60(5). doi:10.5152/j.aott.2026.261021

BibTeX
@article{Aydemir2026,
  author = {Aydemir, Selahaddin and Öztürk, Anıl Murat and Karakaşlı, Ahmet and Hüsemoğlu, Reşit Buğra and Çeltik, Mustafa and Süer, Onur and Akçalı, Ömer},
  title = {Length of the fractured level screw has no significant effect on biomechanical stability in short-segment thoracolumbar fixation—in vitro biomechanical study in a sheep model},
  journal = {Acta Orthopaedica et Traumatologica Turcica},
  year = {2026},
  volume = {60},
  number = {5},
  doi = {10.5152/j.aott.2026.261021}
}
Kostenlos starten →