Preclinical Safety Evaluation of the LapBox Power Tissue Containment System Under Simulated Worst-Case Conditions in a Porcine Model

Yuval Ramot iD, Raphael Lioz, Tal Levin-Harrus, Abraham Nyska iD

Surgical Innovation · 2026

DOI: 10.1177/15533506261415989 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Surgical Innovation, 2026. Verfasst von Yuval Ramot und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Dermatology, Hadassah Medical Center, Jerusalem, Israel. Hauptthemen laut MEDLINE: Laparoscopy, Morcellation, Gynecologic Surgical Procedures. Die Arbeit nennt 5 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Ark Surgical Ltd.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background/Need Uncontained power morcellation during laparoscopic gynecologic surgery risks intra-abdominal dissemination of benign or malignant tissue, a significant safety concern highlighted by FDA warnings. This has created a critical need for robust and reliable tissue containment systems that can be easily integrated into surgical workflows to mitigate this risk. Methodology and device description The LapBox Power Tissue Containment System is a single-use device featuring a dual-walled inflatable chamber designed to create a secure environment for morcellation. We conducted a Good Laboratory Practice (GLP)-compliant toxicology study in three female domestic pigs to assess its safety under simulated worst-case conditions. The device was inserted laparoscopically, and the internal dual-walled chamber of the device was inflated to a high pressure (∼160 mmHg) to simulate a localized worst-case compressive scenario, while the overall intra-abdominal insufflation was maintained at a standard 15 mmHg. Postoperative monitoring included clinical observation, bloodwork, and, at day 13, necropsy and histopathology. Preliminary Results All procedures were completed without mortality, morbidity, or device-related complications. The LapBox maintained full structural integrity. Postoperative clinical, hematological, and biochemical parameters showed no adverse effects. Gross necropsy and detailed histopathology confirmed the absence of device-related ischemia, necrosis, thrombosis, or foreign-body reaction. Current status This preclinical study demonstrates that the LapBox Power system has an excellent safety profile and biocompatibility, even under extreme conditions. The device is ready for the next stage of evaluation. These findings support its translational potential and warrant further investigation in human clinical studies to confirm its safety and efficacy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LaparoscopyMorcellationGynecologic Surgical ProceduresAnimalsFemaleSwineEquipment DesignModels, Animal

Stichwörter der Autor:innen

biomedical engineeringergonomics and/or human factorgynecologic laparoscopy

Autor:innen und Institutionen

  • Department of Dermatology, Hadassah Medical Center, Jerusalem, Israel.
  • The Faculty of Medicine, Hebrew University of Jerusalem, Jerusalem, Israel.
  • HBI Biotech Sciences, a Contract Research GLP-accredited Organization,Weizmann Science Park, Ness Ziona, Israel.
  • Consultant in Toxicologic Pathology, Tel Aviv and Tel Aviv University, Tel Aviv, Israel.

ORCID-Profile: Yuval Ramot, Abraham Nyska

Förderung

  • Ark Surgical Ltd

Zitieren

APA

Ramot, Y., Lioz, R., Levin-Harrus, T., & Nyska, A. (2026). Preclinical Safety Evaluation of the LapBox Power Tissue Containment System Under Simulated Worst-Case Conditions in a Porcine Model. Surgical Innovation. https://doi.org/10.1177/15533506261415989

Vancouver

Ramot Y, Lioz R, Levin-Harrus T, Nyska A. Preclinical Safety Evaluation of the LapBox Power Tissue Containment System Under Simulated Worst-Case Conditions in a Porcine Model. Surgical Innovation. 2026. doi:10.1177/15533506261415989

BibTeX
@article{Ramot2026,
  author = {Ramot, Yuval and Lioz, Raphael and Levin-Harrus, Tal and Nyska, Abraham},
  title = {Preclinical Safety Evaluation of the LapBox Power Tissue Containment System Under Simulated Worst-Case Conditions in a Porcine Model},
  journal = {Surgical Innovation},
  year = {2026},
  doi = {10.1177/15533506261415989}
}
Kostenlos starten →