Validation of a Novel “Windshield Wiper” for Laparoscopes in Cadaver and Live Porcine Models

Christopher R Idelson, Etse-Oghena Y Campbell iD, Lee M Fuentes, Simin E Golestani, Jawad T Ali iD, Austin Fagerberg, Aaron Laviana, John M Uecker

Surgical Innovation · 2025

DOI: 10.1177/15533506251329635 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Surgical Innovation, 2025. Verfasst von Christopher R Idelson und 7 weiteren Autor:innen. Beteiligte Einrichtung: ClearCam Inc, Austin TX, USA. Hauptthemen laut MEDLINE: Laparoscopes, Laparoscopy. Die Arbeit nennt 23 Literaturangaben. Gefördert von: National Institute of Biomedical Imaging and Bioengineering.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

During laparoscopic surgeries, laparoscopes are inserted through a trocar port into the body cavity, which is then insufflated with carbon dioxide. Laparoscope lens clarity frequently becomes compromised via condensation or smearing of blood and adipose. This problem is well-known in the field, yet a viable in vivo solution has yet to address the issue and be successfully clinically adopted. Research Design A structured cadaveric study evaluated the cleaning performance and clinician satisfaction with a novel laparoscope lens cleaning device against 2 gold-standard lens cleaning products. The novel device was also tested in a live animal porcine model to assess cleaning performance in a warm body environment qualitatively. The validation in the porcine model did not have the same evaluation process and comparison with other lens clearing methods as the cadaveric experiment.

Results

Cleaning events were timed individually and analyzed post hoc. Average times to clean scopes for the novel device, Clearify™, and Fred™ Anti-Fog solution were 5 ± 5, 16 ± 7, and 14 ± 6 seconds, respectively. In 100 cleaning events with the novel device, the laparoscope was removed from the body zero times, with an average of 2 ± 1.29 cleaning actuations per event. Clearify™ and Fred™ Anti-Fog were removed from the body 102 and 116 times, with an average number of cleaning actuations of 1.07 ± 0.26 and 1.19 ± 0.53 per event, respectively. In the live porcine model, the novel device consistently cleared all debris deposited on the lens, including fog, tissue, blood, and bile fluid.

Conclusion

This study demonstrates the novel device’s reduction in cleaning duration and scope removals compared to gold-standard technologies, suggesting a potential for improved workflow and reduced intra-operative interruptions.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LaparoscopesLaparoscopyAnimalsSwineCadaverEquipment DesignModels, AnimalHumans

Stichwörter der Autor:innen

laparoscopylensminimally invasive surgery. endoscopyocclusionoperating roomsafetyvisualization

Autor:innen und Institutionen

  • ClearCam Inc, Austin TX, USA.
  • Department of Surgery and Perioperative Care, Dell Medical School, The University of Texas at Austin, Austin TX, USA.
  • McGovern Medical School at the University of Texas Health Science Center, Houston, TX, USA.

ORCID-Profile: Etse-Oghena Y Campbell, Jawad T Ali

Förderung

  • National Institute of Biomedical Imaging and Bioengineering

Zitieren

APA

Idelson, C. R., Campbell, E. Y., Fuentes, L. M., Golestani, S. E., Ali, J. T., Fagerberg, A., Laviana, A., & Uecker, J. M. (2025). Validation of a Novel “Windshield Wiper” for Laparoscopes in Cadaver and Live Porcine Models. Surgical Innovation. https://doi.org/10.1177/15533506251329635

Vancouver

Idelson CR, Campbell EY, Fuentes LM, Golestani SE, Ali JT, Fagerberg A, et al. Validation of a Novel “Windshield Wiper” for Laparoscopes in Cadaver and Live Porcine Models. Surgical Innovation. 2025. doi:10.1177/15533506251329635

BibTeX
@article{Idelson2025,
  author = {Idelson, Christopher R and Campbell, Etse-Oghena Y and Fuentes, Lee M and Golestani, Simin E and Ali, Jawad T and Fagerberg, Austin and Laviana, Aaron and Uecker, John M},
  title = {Validation of a Novel “Windshield Wiper” for Laparoscopes in Cadaver and Live Porcine Models},
  journal = {Surgical Innovation},
  year = {2025},
  doi = {10.1177/15533506251329635}
}
Kostenlos starten →