EUVIMEDEUVIMED entdecken
Multizentrische StudieOpen Access

Post-Market Clinical Follow-Up Study of the LevaLap 1.0 during Laparoscopic Access

Xavier Deffieux, Jean-Louis Benifla, Ingolf Juhasz-Böss, Moritz Breitbach, Olaf Buchweitz, Nassir Habib, Kirsten Hald, Bashar Haj Hamoud, Cyrille Huchon, Vibeke Lysdal, Meletios P Nigdelis, Martin Rudnicki und 6 weitere

JSLS : Journal of the Society of Laparoscopic & Robotic Surgeons · 2025

DOI: 10.4293/jsls.2025.00014 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift JSLS : Journal of the Society of Laparoscopic & Robotic Surgeons, 2025. Verfasst von Xavier Deffieux und 17 weiteren Autor:innen. Beteiligte Einrichtung: Service de Gynécologie, Antoine-Béclère Hospital, Clamart, France. (Dr. Deffieux). Hauptthemen laut MEDLINE: Gynecologic Surgical Procedures, Insufflation, Laparoscopy, Product Surveillance, Postmarketing. Die Arbeit wurde bisher 4-mal zitiert (Crossref). Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Objectives: The LevaLap 1.0 (Core Access Surgical Technologies, Atlanta, GA) was designed to promote safer, more stable, and more predictable abdominal access when using the Veress needle for insufflation. We report on the first postmarket clinical study (PMCF) assessing experience with the use of the LevaLap 1.0 during gynecologic laparoscopic surgery.

Methods

Prospective multicenter study, including women ≥18 years old, excluding pregnancy, access site surgery in prior 10 days, abdominal hernia, contraindication to Veress needle or laparoscopy use, body mass index (BMI) >30 kg/m2, and inability/unwillingness to provide consent.

Results

A total of 158 subjects were included, involving nine surgeons. Mean age was 43.6 ± 14.6 years and mean BMI: 24.7 ± 3.8 kg/m2. Access site was 83.5% transumbilical, 15.2% periumbilical, and 1.2% other. Using the device 96.8% (95% confidence interval [CI]: 92.8-99.0%, n = 152/157) of patients' access was successfully achieved at first attempt and 99.4% (95% CI: 96.5-100.0, n = 156/157) within the first 2 attempts. One minor device-related adverse event was reported: a circular redness on the skin at the site of device application, resolving spontaneously. Surgeons noted easier access in 59.5%, increased confidence in 68.3%, increased access control in 67.1%, and increased access efficiency in 66.4% of cases.

Conclusion

This PMCF study indicates that the use of the LevaLap 1.0 resulted in easier, greater control and greater efficiency during abdominal access using the Veress needle. In 96.8% access was achieved at first attempt and in 99.4% within the first 2 attempts. The use of the LevaLap 1.0 may facilitate abdominal access when using the Veress needle for insufflation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Gynecologic Surgical ProceduresInsufflationLaparoscopyProduct Surveillance, PostmarketingAdultFemaleHumansMiddle AgedEquipment DesignFollow-Up StudiesProspective Studies

Stichwörter der Autor:innen

Abdominal accessLaparoscopyPostmarket clinical follow-upSurgical complicationsVeress needle

Autor:innen und Institutionen

  • Service de Gynécologie, Antoine-Béclère Hospital, Clamart, France. (Dr. Deffieux).
  • Department of Gynecology and Obstetrics, Lariboisière Hospital, Paris, France. (Drs. Benifla and Huchon).
  • Department of Obstetrics and Gynaecology at the Medical Center, University Hospital of Freiburg, Freiburg im Breisgau, Germany (Drs. Juhasz-Böss, Moritz Breitbach, and Taran).
  • Frauenklinik an der Elbe, Hamburg, Germany. (Drs. Buchweitz and Hackethal).
  • Clinique de l'Yvette, Longjumeau, France. (Dr. Habib).
  • Division of Gynecology and Obstetrics, Oslo University Hospital, Oslo, Norway. (Drs. Hald and Michelsen-Wahl).

Zitieren

APA

Deffieux, X., Benifla, J. L., Juhasz-Böss, I., Breitbach, M., Buchweitz, O., Habib, N., Hald, K., Haj Hamoud, B., Huchon, C., Lysdal, V., Nigdelis, M. P., Rudnicki, M., Solomayer, E. F., Taran, F. A., Michelsen-Wahl, H., Azziz, R., Bagnardi, V., & Hackethal, A. (2025). Post-Market Clinical Follow-Up Study of the LevaLap 1.0 during Laparoscopic Access. JSLS : Journal of the Society of Laparoscopic & Robotic Surgeons. https://doi.org/10.4293/jsls.2025.00014

Vancouver

Deffieux X, Benifla JL, Juhasz-Böss I, Breitbach M, Buchweitz O, Habib N, et al. Post-Market Clinical Follow-Up Study of the LevaLap 1.0 during Laparoscopic Access. JSLS : Journal of the Society of Laparoscopic & Robotic Surgeons. 2025. doi:10.4293/jsls.2025.00014

BibTeX
@article{Deffieux2025,
  author = {Deffieux, Xavier and Benifla, Jean-Louis and Juhasz-Böss, Ingolf and Breitbach, Moritz and Buchweitz, Olaf and Habib, Nassir and Hald, Kirsten and Haj Hamoud, Bashar and Huchon, Cyrille and Lysdal, Vibeke and Nigdelis, Meletios P and Rudnicki, Martin and Solomayer, Erich-Franz and Taran, Florin-Andrei and Michelsen-Wahl, Henrik and Azziz, Ricardo and Bagnardi, Vincenzo and Hackethal, Andreas},
  title = {Post-Market Clinical Follow-Up Study of the LevaLap 1.0 during Laparoscopic Access},
  journal = {JSLS : Journal of the Society of Laparoscopic & Robotic Surgeons},
  year = {2025},
  doi = {10.4293/jsls.2025.00014}
}
Kostenlos starten →