Proton Beam Range and Charge Verification Using Multilayer Faraday Collector
Technology in Cancer Research & Treatment · 2024
DOI: 10.1177/15330338241262610 ↗Kurz gefasst
Fachartikel in der Zeitschrift Technology in Cancer Research & Treatment, 2024. Verfasst von Ping L Yeap und 8 weiteren Autor:innen. Beteiligte Einrichtung: Division of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore. Hauptthemen laut MEDLINE: Proton Therapy. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Duke-NUS Oncology Academic Program, Clinical & Systems Innovation Support – Innovation Seed Grant.
Abstract (Originalsprache)
Purpose
A daily quality assurance (QA) check in proton therapy is ensuring that the range of each proton beam energy in water is accurate to 1 mm. This is important for ensuring that the tumor is adequately irradiated while minimizing damage to surrounding healthy tissue. It is also important to verify the total charge collected against the beam model. This work proposes a time-efficient method for verifying the range and total charge of proton beams at different energies using a multilayer Faraday collector (MLFC).
Methods
We used an MLFC-128-250 MeV comprising 128 layers of thin copper foils separated by thin insulating Kapton TM layers. Protons passing through the collector induce a charge on the metallic foils, which is integrated and measured by a multichannel electrometer. The charge deposition on the foils provides information about the beam range.
Results
Our results show that the proton beam range obtained using MLFC correlates closely with the range obtained from commissioning water tank measurements for all proton energies. Upon applying a range calibration factor, the maximum deviation is 0.4 g/cm 2 . The MLFC range showed no dependence on the number of monitor units and the source-to-surface distance. Range measurements collected over multiple weeks exhibited stability. The total charge collected agrees closely with the theoretical charge from the treatment planning system beam model for low- and mid-range energies.
Conclusions
We have calibrated and commissioned the use of the MLFC to easily verify range and total charge of proton beams. This tool will improve the workflow efficiency of the proton QA.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Division of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore.
- Department of Oncology, University of Cambridge, UK.
- Division of Physics and Applied Physics, Nanyang Technological University, Singapore, Singapore.
- Oncology Academic Clinical Programme, Duke-NUS Medical School, Singapore, Singapore.
ORCID-Profile: Ping L Yeap, Wei Y C Koh
Förderung
- Duke-NUS Oncology Academic Program
- Clinical & Systems Innovation Support – Innovation Seed Grant
Zitieren
APA
Yeap, P. L., Lew, K. S., Koh, W. Y. C., Chua, C. G. A., Wibawa, A., Master, Z., Lee, J. C. L., Park, S. Y., & Tan, H. Q. (2024). Proton Beam Range and Charge Verification Using Multilayer Faraday Collector. Technology in Cancer Research & Treatment. https://doi.org/10.1177/15330338241262610
Vancouver
Yeap PL, Lew KS, Koh WYC, Chua CGA, Wibawa A, Master Z, et al. Proton Beam Range and Charge Verification Using Multilayer Faraday Collector. Technology in Cancer Research & Treatment. 2024. doi:10.1177/15330338241262610
BibTeX
@article{Yeap2024,
author = {Yeap, Ping L and Lew, Kah S and Koh, Wei Y C and Chua, Clifford G A and Wibawa, Andrew and Master, Zubin and Lee, James C L and Park, Sung Y and Tan, Hong Q},
title = {Proton Beam Range and Charge Verification Using Multilayer Faraday Collector},
journal = {Technology in Cancer Research & Treatment},
year = {2024},
doi = {10.1177/15330338241262610}
}
EUVIMED