Effect of different sealing caps on the backflow of short-term dialysis catheters: A bench study
The Journal of Vascular Access · 2025
DOI: 10.1177/11297298241301508 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift The Journal of Vascular Access, 2025. Verfasst von Daniele Privitera und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedicine and Prevention, University of Rome Tor Vergata, Rome, Italy. Hauptthemen laut MEDLINE: Renal Dialysis, Catheters, Indwelling, Catheterization, Central Venous, Central Venous Catheters. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 24 Literaturangaben.
Abstract (Originalsprache)
Background
Needle-free connectors (NFCs) are recommended as closure systems for peripheral and central vascular catheters to reduce needlestick injuries and infections, while potentially reducing blood reflux. However, their performance in short-term dialysis catheters has never been evaluated. The aim of this study was to evaluate the backflow associated with two NFCs (Neutron™ and Tego™) compared to the standard closure.
Methods
In this bench study, the physiological blood pressure of the superior vena cava was simulated. Expert operators performed three sealing sequences for each combination of connector and lumens of the catheter, with and without closing the clamp. After that, the backflow was measured in millimeters using a high-precision digital caliper.
Results
No combination of caps, lumens, or clamping sequences eliminated the backflow. No differences were observed between standard caps and NFCs in both lumens, apart from the Tego™/ No clamp combination in the proximal lumen that showed higher backflow (standard cap 15 [11; 17] mm vs Tego™ 23 [19; 25] mm, p < 0.001). Clamping reduced backflow in both the proximal (13 [11; 17] mm vs 20 [13; 25] mm) and distal lumens (12 [11; 16] mm vs 14 [12; 17] mm). No differences were found between standard caps and NFCs in the distal lumen, regardless of clamping. Proximal lumen consistently exhibited larger backflow (14 [12; 22] mm) compared to distal lumens (13 [11; 17] mm, p = 0.005).
Conclusion
NFCs do not seem to introduce any advantage on backflow reduction as compared to a standard cap combined with lumen and clamping. Our data suggest that clamping should become standard practice when NFCs are used in the management of short-term dialysis catheters.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedicine and Prevention, University of Rome Tor Vergata, Rome, Italy.
- Department of Cardiovascular, Neural and Metabolic Sciences, Istituto Auxologico Italiano, IRCCS, San Luca Hospital, Milan, Italy.
- Post Anesthesia Care Unit, Vascular Access Team ASST Lodi, Lodi, Italy.
- Department of Medicine and Surgery, University of Milan-Bicocca, Monza, Italy.
- Department of Nephrology, University Maggiore della Carità Hospital, Novara, Italy.
- Department of Emergency and Intensive Care, University Maggiore della Carità Hospital, Novara, Italy.
ORCID-Profile: Davide Giustivi
Zitieren
APA
Privitera, D., Giustivi, D., Langer, T., Fiorina, E., Gotti, F., Rossini, M., Brunoni, B., Capsoni, N., Dal Molin, A., & Zadek, F. (2025). Effect of different sealing caps on the backflow of short-term dialysis catheters: A bench study. The Journal of Vascular Access. https://doi.org/10.1177/11297298241301508
Vancouver
Privitera D, Giustivi D, Langer T, Fiorina E, Gotti F, Rossini M, et al. Effect of different sealing caps on the backflow of short-term dialysis catheters: A bench study. The Journal of Vascular Access. 2025. doi:10.1177/11297298241301508
BibTeX
@article{Privitera2025,
author = {Privitera, Daniele and Giustivi, Davide and Langer, Thomas and Fiorina, Elisabetta and Gotti, Federica and Rossini, Michela and Brunoni, Beatrice and Capsoni, Nicolò and Dal Molin, Alberto and Zadek, Francesco},
title = {Effect of different sealing caps on the backflow of short-term dialysis catheters: A bench study},
journal = {The Journal of Vascular Access},
year = {2025},
doi = {10.1177/11297298241301508}
}
EUVIMED