EUVIMEDEUVIMED entdecken
Fall-Kontroll-Studie

Effects of indwelling centrally inserted central catheter on tip location of peripherally inserted central catheter with intracavitary electrocardiogram guidance: A retrospective case-control study

Wenyan Sun, Jiaqian Li iD, Bing Liu, Yue Liu, Ruibin Ge, Kai Wang, Jingyan Zhang, Xinjuan Wu iD

The Journal of Vascular Access · 2023-05 · 24 (3) · 379-384

DOI: 10.1177/11297298211015088 ↗

Kurz gefasst

Fall-Kontroll-Studie in der Zeitschrift The Journal of Vascular Access, online erschienen am 23.07.2021. Verfasst von Wenyan Sun und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Nutrition, Peking Union Medical College Hospital, Beijing, China. Hauptthemen laut MEDLINE: Catheterization, Central Venous, Central Venous Catheters, Catheterization, Peripheral. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 28 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Intracavity electrocardiogram (IC-ECG) guidance is an effective technology for monitoring the tip locations of centrally inserted central catheters (CICCs) and peripherally inserted central catheters (PICCs). However, for patients with an indwelling CICC, there has been no study on whether the CICC affects PICC tip positioning with IC-ECG guidance.

Methods

Thirty people with a CICC requiring PICC insertion, matched with age to controls without a CICC. The stability of IC-ECG waveforms, the amplitude of P-waves, and the accuracy of PICC tip location were compared.

Results

There was no significant difference in the stability of the IC-ECG waveforms (χ 2 = 0.22, p = 0.64). The amplitudes of baseline P-waves and ideal P-waves also showed no significant difference ( t1 = 0.06, p1 = 0.96, t2 = 0.80, p2 = 0.43). Neither the accuracy of tip location (χ 2 = 1.40, p = 0.50) nor the distance of PICC tip ( t = −0.03, p = 0.98) were significantly different.

Conclusion

For patients with an indwelling CICC, the position of PICC tip can be accurately determined by the dynamic changes in the P-wave amplitude.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Catheterization, Central VenousCentral Venous CathetersCatheterization, PeripheralHumansRetrospective StudiesCase-Control StudiesElectrocardiography

Stichwörter der Autor:innen

Peripherally inserted central cathetercentrally inserted central catheterintracavitary electrocardiogramtip location

Autor:innen und Institutionen

  • Department of Clinical Nutrition, Peking Union Medical College Hospital, Beijing, China.
  • Department of Nursing, Peking Union Medical College Hospital, Beijing, China.

ORCID-Profile: Jiaqian Li, Xinjuan Wu

Zitieren

APA

Sun, W., Li, J., Liu, B., Liu, Y., Ge, R., Wang, K., Zhang, J., & Wu, X. (2023). Effects of indwelling centrally inserted central catheter on tip location of peripherally inserted central catheter with intracavitary electrocardiogram guidance: A retrospective case-control study. The Journal of Vascular Access, 24(3), 379-384. https://doi.org/10.1177/11297298211015088

Vancouver

Sun W, Li J, Liu B, Liu Y, Ge R, Wang K, et al. Effects of indwelling centrally inserted central catheter on tip location of peripherally inserted central catheter with intracavitary electrocardiogram guidance: A retrospective case-control study. The Journal of Vascular Access. 2023;24(3):379-384. doi:10.1177/11297298211015088

BibTeX
@article{Sun2023,
  author = {Sun, Wenyan and Li, Jiaqian and Liu, Bing and Liu, Yue and Ge, Ruibin and Wang, Kai and Zhang, Jingyan and Wu, Xinjuan},
  title = {Effects of indwelling centrally inserted central catheter on tip location of peripherally inserted central catheter with intracavitary electrocardiogram guidance: A retrospective case-control study},
  journal = {The Journal of Vascular Access},
  year = {2023},
  volume = {24},
  number = {3},
  pages = {379--384},
  doi = {10.1177/11297298211015088}
}
Kostenlos starten →