Impact of distance of the catheter tip from cavo-atrial junction on bubble test (delay) time: A prospective study
The Journal of Vascular Access · 2024
DOI: 10.1177/11297298231153517 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Journal of Vascular Access, 2024. Verfasst von Elli Stefano und 11 weiteren Autor:innen. Beteiligte Einrichtung: University of Milan-Bicocca, A.S.S.T. Monza, San Gerardo Hospital, Monza (MB), Italy. Hauptthemen laut MEDLINE: Catheterization, Central Venous, Central Venous Catheters, Predictive Value of Tests, Catheters, Indwelling. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 23 Literaturangaben.
Abstract (Originalsprache)
Introduction
Correct tip positioning is a critical aspect in central vascular access devices insertion. The verification of positioning at the cavo-atrial junction is usually performed by intracavitary electrocardiography. Recently, echocardiographic techniques were proposed, including the direct visualization of the catheter or the visualization of a saline/air bolus (i.e. “bubble test”). As for the latter, a push-to-bubbles delay time below 2 s was proposed to indicate a correct positioning of the catheter tip. The aim of this study was to measure the variations of the push-to-bubbles time at increasing distance from the cavo-atrial junction, to verify if a cut-off of 1–2 s correspond to a well-positioned catheter.
Methods
We performed a prospective study including patients with clinical indication of positioning a peripherally inserted central catheter. The catheter tip was positioned at the cavo-atrial junction (P0) via intracavitary electrocardiography, and the push-to-bubbles delay time was measured. The catheter was then retracted 5 cm (P1) and 10 cm (P2), and the test was repeated at this positioning. Push-to-bubbles time measurements were performed off-line by analyzing an audio/video recording which included the echography screen and the voice signal of the operator.
Results
Forty-nine patients were included. The average push-to-bubble time when the catheter tip was in the reference position was 0.41 ± 0.21 s. Retraction of the PICC catheter of 5 and 10 cm determined a significant increase of the push-to-bubbles time: mean time difference was +0.34 (95% IC 0.25–0.43, p < 0.001) s between P0 and P1 (5 cm distance), and +0.77 (95% IC 0.62–0.92, p < 0.001) s between P0 and P2 (10 cm distance). When the catheter was at the reference position (i.e. cavo-atrial junction) only 2.1% of bubbles delay times were above 1 s.
Conclusion
The push-to-bubbles time is very low when the catheter tip is at the cavo-atrial junction. This delay increases progressively with increasing distance from the target. Push-to-bubbles delay time above 1 s might indicate a catheter not close to the cavo-atrial junction.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University of Milan-Bicocca, A.S.S.T. Monza, San Gerardo Hospital, Monza (MB), Italy.
- A.S.S.T. Monza, San Gerardo Hospital, Monza (MB), Italy.
- ASST-Lodi Codogno Hospital, Codogno, Italy.
- ASST-Lariana, Italy.
- IRCCS Foundation Maggiore Policlinico Hospital, Milan, Italy.
- ASST-Brianza, Vimercate Hospital, Vimercate, Italy.
ORCID-Profile: Elli Stefano, Lucchini Alberto, Giani Marco
Zitieren
APA
Stefano, E., Dario, D., Silvia, C., Gloria, C., Mariavittoria, G., Francesco, M., Jessica, P., Mario, R., Alberto, L., Giuseppe, F., Matteo, P., & Marco, G. (2024). Impact of distance of the catheter tip from cavo-atrial junction on bubble test (delay) time: A prospective study. The Journal of Vascular Access. https://doi.org/10.1177/11297298231153517
Vancouver
Stefano E, Dario D, Silvia C, Gloria C, Mariavittoria G, Francesco M, et al. Impact of distance of the catheter tip from cavo-atrial junction on bubble test (delay) time: A prospective study. The Journal of Vascular Access. 2024. doi:10.1177/11297298231153517
BibTeX
@article{Stefano2024,
author = {Stefano, Elli and Dario, D'amata and Silvia, Cavalli and Gloria, Cambiaghi and Mariavittoria, Giorgianni and Francesco, Marino and Jessica, Pozzoli and Mario, Russo and Alberto, Lucchini and Giuseppe, Foti and Matteo, Pozzi and Marco, Giani},
title = {Impact of distance of the catheter tip from cavo-atrial junction on bubble test (delay) time: A prospective study},
journal = {The Journal of Vascular Access},
year = {2024},
doi = {10.1177/11297298231153517}
}
EUVIMED