Randomisierte kontrollierte Studie in HIP International, 2024. Written by Amanda D Klaassen and 8 co-authors. Institution: Department of Orthopaedic Surgery, OLVG, Amsterdam, the Netherlands. Major MEDLINE topics: Arthroplasty, Replacement, Hip, Hip Prosthesis, Radiostereometric Analysis, Ceramics. 28 references. The full text is freely available (open access, CC BY 4.0).
Abstract (original language)
Introduction
The inelasticity of ceramic bearings might affect primary stability and migration of implants in press-fit total hip arthroplasty (THA). This randomised controlled trial compares migration patterns of the uncemented Delta-TT cup and H-MAX S stem between ceramic and polyethylene liners, up to 5 years follow-up.
Methods
Patients receiving primary press-fit THA were randomly allocated to a ceramic ( n = 28) or polyethylene ( n = 25) liner. Migration was measured using model-based radiostereometric analysis (RSA) at baseline and 1.5, 3, 6, 12, 24 and 60 months postoperatively and compared between groups using mixed models statistical analysis. The focus of this study is on the 2- to 5-year migration of the Delta-TT cup and migration during complete follow-up of the H-MAX S stem up to 5-years.
Results
At 5-year follow-up, mean (95% CI) proximal cup translation was 0.56 (0.37–0.74) mm in ceramic and 0.58 (0.25–0.90) mm in polyethylene ( p = 0.729). The mean change in adduction was 1.05° (0.27–1.82°) in ceramic and 0.78° (−0.16–1.71°) in polyethylene. Mixed modelling showed that all between-group effects were ⩽0.20 mm for translation and ⩽0.22° for rotation at 5 years postoperatively ( p ⩾ 0.23). Mean cup migration between 2 and 5 years was limited (all parameters
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Department of Orthopaedic Surgery, OLVG, Amsterdam, the Netherlands.
- Department of Orthopaedic Surgery, LUMC, Leiden, the Netherlands.
- Xpert Clinics Orthopedie Amsterdam, the Netherlands.
- Department of Orthopaedic Surgery, Amsterdam Movement Sciences, Amsterdam UMC, Academic Medical Centre, University of Amsterdam, Amsterdam, the Netherlands.
- Department of Radiology, OLVG, Amsterdam, the Netherlands.
ORCID profiles: Amanda D Klaassen, Justin van Loon
Cite
APA
Klaassen, A. D., van Loon, J., Willigenburg, N. W., Koster, L. A., Kaptein, B. L., van der Hulst, V. P. M., Haverkamp, D., Moojen, D. J. F., & Poolman, R. W. (2024). Comparison of 5-year cup and stem migration between ceramic-on-ceramic and ceramic-on-polyethylene bearing in press-fit total hip arthroplasty: a randomised controlled trial using radiostereometric analysis. HIP International. https://doi.org/10.1177/11207000241265653
Vancouver
Klaassen AD, van Loon J, Willigenburg NW, Koster LA, Kaptein BL, van der Hulst VPM, et al. Comparison of 5-year cup and stem migration between ceramic-on-ceramic and ceramic-on-polyethylene bearing in press-fit total hip arthroplasty: a randomised controlled trial using radiostereometric analysis. HIP International. 2024. doi:10.1177/11207000241265653
BibTeX
@article{Klaassen2024,
author = {Klaassen, Amanda D and van Loon, Justin and Willigenburg, Nienke W and Koster, Lennard A and Kaptein, Bart L and van der Hulst, Victor P M and Haverkamp, Daniel and Moojen, Dirk Jan F and Poolman, Rudolf W},
title = {Comparison of 5-year cup and stem migration between ceramic-on-ceramic and ceramic-on-polyethylene bearing in press-fit total hip arthroplasty: a randomised controlled trial using radiostereometric analysis},
journal = {HIP International},
year = {2024},
doi = {10.1177/11207000241265653}
}
EUVIMED