Development and preliminary evaluation of a transducer guiding system for hand ultrasound: With a focus on reproducibility and operational stability

Mu-Ting Wang, Yu-Chen Lin, Chieh-Hsiang Hsu, Hsiu-Yun Hsu, Tai-Hua Yang, Li-Chieh Kuo iD, Fong-Chin Su

Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine · 2026

DOI: 10.1177/09544119251414217 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2026. Verfasst von Mu-Ting Wang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, College of Engineering, National Cheng Kung University, Tainan City, Taiwan. Hauptthemen laut MEDLINE: Transducers, Hand. Die Arbeit nennt 55 Literaturangaben. Gefördert von: National Science and Technology Council, E-Da Hospital, national cheng kung university.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Various strategies have been proposed to reduce operator-dependent variability in musculoskeletal ultrasound, including mechanical stabilization techniques. However, their effects on image reproducibility and transducer handling remain unclear in anatomically complex regions such as hand. This two-phase study aimed to (1) develop and validate a transducer guiding system for hand ultrasound and (2) evaluate its influence on image reproducibility and operational stability through image-based analyses. In Phase I, test–retest reliability of the guiding system was examined by repeatedly measuring soft-tissue thickness in 30 healthy participants. Intraclass Correlation Coefficients (ICC) ranged from 0.766 to 0.948, demonstrating good to excellent reliability. In Phase II, 16 ultrasound users acquired images under handheld and guiding-system-assisted conditions based on predefined reference images. Image reproducibility was evaluated using Normalized Cross-Correlation (NCC) and ICC, while operational stability was assessed via cosine similarity derived from M-mode segments during the pre-capture period, with group comparisons performed using the Wilcoxon signed-rank test. Novice users showed substantial ICC improvement when using the guiding system (0.487–0.681), approaching the consistency observed in experienced users. Both novice and experienced groups displayed higher NCC and cosine similarity values with the guiding system, indicating improved reproducibility and operational stability. Overall, the guiding system enhanced image outcomes by providing standardized positioning and mechanical stabilization. These findings highlight its potential to improve consistency in hand ultrasound assessments and serve as a supportive tool for novice ultrasound training. The results further demonstrate the feasibility of mechanical stabilization in reducing operator-related variability and enhancing image consistency in hand ultrasound examinations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

TransducersHandHumansUltrasonographyReproducibility of ResultsMaleEquipment DesignImage Processing, Computer-AssistedFemaleAdult

Stichwörter der Autor:innen

guiding systemhandimage reproducibilitymusculoskeletal ultrasoundoperational stabilityrepeated measurements

Autor:innen und Institutionen

  • Department of Biomedical Engineering, College of Engineering, National Cheng Kung University, Tainan City, Taiwan.
  • Department of Occupational Therapy, Da-Yeh University, Changhua, Taiwan.
  • Department of Mechanical Engineering, Southern Taiwan University of Science and Technology, Tainan City, Taiwan.
  • Department of Occupational Therapy, College of Medicine, National Cheng Kung University, Tainan City, Taiwan.
  • Department of Physical Medicine and Rehabilitation, National Cheng Kung University Hospital, College of Medicine, National Cheng Kung University, Tainan City, Taiwan.
  • Medical Device Innovation Center, National Cheng Kung University, Tainan City, Taiwan.

ORCID-Profile: Li-Chieh Kuo

Förderung

  • National Science and Technology Council
  • E-Da Hospital
  • national cheng kung university
  • ministry of education

Zitieren

APA

Wang, M. T., Lin, Y. C., Hsu, C. H., Hsu, H. Y., Yang, T. H., Kuo, L. C., & Su, F. C. (2026). Development and preliminary evaluation of a transducer guiding system for hand ultrasound: With a focus on reproducibility and operational stability. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. https://doi.org/10.1177/09544119251414217

Vancouver

Wang MT, Lin YC, Hsu CH, Hsu HY, Yang TH, Kuo LC, et al. Development and preliminary evaluation of a transducer guiding system for hand ultrasound: With a focus on reproducibility and operational stability. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 2026. doi:10.1177/09544119251414217

BibTeX
@article{Wang2026,
  author = {Wang, Mu-Ting and Lin, Yu-Chen and Hsu, Chieh-Hsiang and Hsu, Hsiu-Yun and Yang, Tai-Hua and Kuo, Li-Chieh and Su, Fong-Chin},
  title = {Development and preliminary evaluation of a transducer guiding system for hand ultrasound: With a focus on reproducibility and operational stability},
  journal = {Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine},
  year = {2026},
  doi = {10.1177/09544119251414217}
}