Deep learning based time-dependent reliability analysis of an underactuated lower-limb robot exoskeleton for gait rehabilitation
Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine · 2025
DOI: 10.1177/09544119251349362 ↗Kurz gefasst
Fachartikel in der Zeitschrift Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2025. Verfasst von Fahad Hussain und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Information Technology and Systems, University of Canberra, Canberra, ACT, Australia. Hauptthemen laut MEDLINE: Deep Learning, Exoskeleton Device, Gait, Lower Extremity. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben.
Abstract (Originalsprache)
This study evaluates the reliability of an underactuated wearable lower-limb exoskeleton designed to assist with gait rehabilitation. Recognizing the complexity of system reliability, a deep learning framework augmented with Long short-term Memory (LSTM) was utilized for the time-dependent reliability analysis of dynamic systems. The research commenced with the development of a lower-limb gait robot, modeled on a Stephenson III six-bar linkage mechanism. Following the mechanical design, computer-aided design (CAD) tools were employed to conceptualize a lower-limb robotic exoskeleton for rehabilitation purposes. The design incorporated two metallic materials (aluminum and steel), and a composite material (carbon fiber) tested using SolidWorks ® . The prototype achieved a lightweight design (~1.63 kg) for carbon fiber material. An LSTM-enhanced deep neural network algorithm was implemented to predict the time-dependent reliability of joint displacements and end-effector trajectories. Finally, conditional probability methods were applied to complete the time-dependent system reliability assessment. The designed mechanical system for gait rehabilitation demonstrated high reliability ( R ≈ 0.87). Over 200 simulation runs, reliability trends showed consistent and robust predictions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Information Technology and Systems, University of Canberra, Canberra, ACT, Australia.
- School of Engineering and Digital Sciences, Nazarbayev University, Astana, Kazakhstan.
- School of Systems and Computing, University of New South Wales, Canberra, ACT, Australia.
ORCID-Profile: Fahad Hussain, Shahid Hussain, Prashant Jamwal
Zitieren
APA
Hussain, F., Goyal, T., Hussain, S., Jamwal, P., & Goecke, R. (2025). Deep learning based time-dependent reliability analysis of an underactuated lower-limb robot exoskeleton for gait rehabilitation. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. https://doi.org/10.1177/09544119251349362
Vancouver
Hussain F, Goyal T, Hussain S, Jamwal P, Goecke R. Deep learning based time-dependent reliability analysis of an underactuated lower-limb robot exoskeleton for gait rehabilitation. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 2025. doi:10.1177/09544119251349362
BibTeX
@article{Hussain2025,
author = {Hussain, Fahad and Goyal, Tanishka and Hussain, Shahid and Jamwal, Prashant and Goecke, Roland},
title = {Deep learning based time-dependent reliability analysis of an underactuated lower-limb robot exoskeleton for gait rehabilitation},
journal = {Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine},
year = {2025},
doi = {10.1177/09544119251349362}
}
EUVIMED