Pulmonary nodule visualization and evaluation of AI-based detection at various ultra-low-dose levels using photon-counting detector CT
Acta Radiologica · 2024-10 · 65 (10) · 1238-1245
DOI: 10.1177/02841851241275289 ↗Kurz gefasst
Fachartikel in der Zeitschrift Acta Radiologica, online erschienen am 15.09.2024. Verfasst von Lisa Jungblut und 6 weiteren Autor:innen. Beteiligte Einrichtung: Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland. Hauptthemen laut MEDLINE: Radiation Dosage, Tomography, X-Ray Computed, Phantoms, Imaging, Solitary Pulmonary Nodule. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 43 Literaturangaben.
Abstract (Originalsprache)
Background
Radiation dose should be as low as reasonably achievable. With the invention of photon-counting detector computed tomography (PCD-CT), the radiation dose may be considerably reduced.
Purpose
To evaluate the potential of PCD-CT for dose reduction in pulmonary nodule visualization for human readers as well as for computer-aided detection (CAD) studies. Material and Methods A chest phantom containing pulmonary nodules of different sizes/densities (range 3–12 mm and −800–100 HU) was scanned on a PCD-CT with standard low-dose protocol as well as with half, quarter, and 1/40 dose (CTDI vol 0.4–0.03 mGy). Dose-matched scans were performed on a third-generation energy-integrating detector CT (EID-CT). Evaluation of nodule visualization and detectability was performed by two blinded radiologists. Subjective image quality was rated on a 5-point Likert scale. Artificial intelligence (AI)-based nodule detection was performed using commercially available software.
Results
Highest image noise was found at the lowest dose setting of 1/40 radiation dose (eff. dose = 0.01mSv) with 166.1 ± 18.5 HU for PCD-CT and 351.8 ± 53.0 HU for EID-CT. Overall sensitivity was 100% versus 93% at standard low-dose protocol (eff. dose = 0.2 mSv) for PCD-CT and EID-CT, respectively. At the half radiation dose, sensitivity remained 100% for human reader and CAD studies in PCD-CT. At the quarter radiation dose, PCD-CT achieved the same results as EID-CT at the standard radiation dose setting (93%, P = 1.00) in human reading studies. The AI-CAD system delivered a sensitivity of 93% at the lowest radiation dose level in PCD-CT.
Conclusion
At half dose, PCD CT showed pulmonary nodules similar to full-dose PCD, and at quarter dose, PCD CT performed comparably to standard low-dose EID CT. The CAD algorithm is effective even at ultra-low doses.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
ORCID-Profile: Lisa Jungblut
Zitieren
APA
Jungblut, L., Euler, A., Landsmann, A., Englmaier, V., Mergen, V., Sefirovic, M., & Frauenfelder, T. (2024). Pulmonary nodule visualization and evaluation of AI-based detection at various ultra-low-dose levels using photon-counting detector CT. Acta Radiologica, 65(10), 1238-1245. https://doi.org/10.1177/02841851241275289
Vancouver
Jungblut L, Euler A, Landsmann A, Englmaier V, Mergen V, Sefirovic M, et al. Pulmonary nodule visualization and evaluation of AI-based detection at various ultra-low-dose levels using photon-counting detector CT. Acta Radiologica. 2024;65(10):1238-1245. doi:10.1177/02841851241275289
BibTeX
@article{Jungblut2024,
author = {Jungblut, Lisa and Euler, André and Landsmann, Anna and Englmaier, Vanessa and Mergen, Victor and Sefirovic, Medina and Frauenfelder, Thomas},
title = {Pulmonary nodule visualization and evaluation of AI-based detection at various ultra-low-dose levels using photon-counting detector CT},
journal = {Acta Radiologica},
year = {2024},
volume = {65},
number = {10},
pages = {1238--1245},
doi = {10.1177/02841851241275289}
}
EUVIMED