The effect of excessive gas to blood ratios in an ECMO oxygenator
Perfusion · 2025
DOI: 10.1177/02676591241256089 ↗Kurz gefasst
Fachartikel in der Zeitschrift Perfusion, 2025. Verfasst von Michael Shaw und 5 weiteren Autor:innen. Beteiligte Einrichtung: Perfusion Department, Great Ormond Street Hospital for Children, London, UK. Hauptthemen laut MEDLINE: Extracorporeal Membrane Oxygenation, Oxygenators, Membrane, Oxygen, Embolism, Air. Die Arbeit nennt 11 Literaturangaben.
Abstract (Originalsprache)
Introduction
Oxygenators for paediatric Extracorporeal Membrane Oxygenation (ECMO) are required to operate over a wide range of flow rates, in a patient group ranging from neonates through to fully grown adolescents. ECMO oxygenators typically have a manufacturer’s stated maximum gas: blood flow rate (GBFR) ratio of 2:1, however, many patients require greater ratios than this for adequate CO 2 removal. Mismatches in GBFR in theory could result in high gas phase pressures. These increased pressures in theory could cause the formation of gross gaseous microemboli (GME) placing the child at higher risk of neurological injury.
Methods
We evaluated 6 paediatric and 6 adult A.L.ONE™ ECMO oxygenators and assessed their gas phase pressures and GME release, in an ex vivo setting, in GBFR ratios up to greater than 2, across a range of gas flow (1L – 10 L/min) rates with a fraction of inspired oxygen (F i O 2 ) content of 50% and 100%.
Results
There were no increases above 10 mmHg observed in gas phase pressures in GBFR >= 2:1 in either adult or paediatric oxygenators. Laboratory examination of GME activity demonstrated a small increase in post-membrane GME release over the study period. GME release was unaffected by F i O 2 setting or gas flow rate, with a maximum volume of < 6 µL in both paediatric and adult oxygenators.
Conclusions
In an ex vivo setting, increasing GBFR above 2:1 in a paediatric oxygenator, and to a GBFR of 2:1 in an adult oxygenator did not significantly increase gas phase pressures, and no oxygenator membrane rupture was observed. There were no associations between gas flow rates and GME production.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Perfusion Department, Great Ormond Street Hospital for Children, London, UK.
- Cardiac Intensive Care Department, Great Ormond Street Hospital, London, UK.
- Centre for Heart Failure, Transplantation and Extracorporeal Support, Research Department of Children's Cardiovascular Disease, UCL Institute of Cardiovascular Science, London, UK.
ORCID-Profile: Richard Crook, Richard W Issitt
Zitieren
APA
Shaw, M., Cross, N., Richardson, R., Crook, R., Thirulchelvam, T., & Issitt, R. W. (2025). The effect of excessive gas to blood ratios in an ECMO oxygenator. Perfusion. https://doi.org/10.1177/02676591241256089
Vancouver
Shaw M, Cross N, Richardson R, Crook R, Thirulchelvam T, Issitt RW. The effect of excessive gas to blood ratios in an ECMO oxygenator. Perfusion. 2025. doi:10.1177/02676591241256089
BibTeX
@article{Shaw2025,
author = {Shaw, Michael and Cross, Nigel and Richardson, Rebecca and Crook, Richard and Thirulchelvam, Timothy and Issitt, Richard W},
title = {The effect of excessive gas to blood ratios in an ECMO oxygenator},
journal = {Perfusion},
year = {2025},
doi = {10.1177/02676591241256089}
}
EUVIMED