Comparative analysis of oxygenator dysfunction in polymethylpentene oxygenators: A pilot study
Perfusion · 2025
DOI: 10.1177/02676591241268402 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Perfusion, 2025. Verfasst von Shan P Modi und 7 weiteren Autor:innen. Beteiligte Einrichtung: Cardiovascular Institute, Allegheny Health Network, Pittsburgh, PA, USA. Hauptthemen laut MEDLINE: Extracorporeal Membrane Oxygenation, Oxygenators, Membrane. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 19 Literaturangaben.
Abstract (Originalsprache)
Introduction
Polymethylpentene (PMP) oxygenators serve as the primary oxygenator type utilized for ECMO. With the number of PMP oxygenators available, it has become increasingly important to determine differences among each oxygenator type that can lead to varying metrics of oxygenator dysfunction.
Methods
This study was a retrospective, single-center analysis of adult patients supported on ECMO between December 2020 to December 2021 with varying PMP oxygenators including the Medtronic Nautilus Smart (Minneapolis, MA), the Eurosets AMG PMP (Medolla, Italy) and Getinge Quadrox-iD and the Getinge Cardiohelp HLS Module Advanced System (Gothenberg, Sweden).
Results
A total of 19 patients were included in our study. 10 patients (52.6%) were supported with a Medtronic Nautilus Smart oxygenator, 5 patients (26.3%) were supported with an Eurosets AMG PMP Oxygenator, and 4 patients (21.1%) were supported with either a Getinge Quadrox-iD oxygenator or Getinge Cardiohelp HLS system. Patients supported with Eurosets AMG PMP oxygenators experienced higher resistance and lower post-oxygenator PaO 2 in comparison to other cohorts ( p < .02 and < .002 respectively). There was no difference in measured oxygen transfer between cohorts ( p = .667). Two patients, both supported by Eurosets AMG PMP, experienced oxygenator failure ( p = . 094).
Conclusion
Radial flow oxygenators are prone to higher resistance and lower post-oxygenator PaO 2 when compared to transverse flow oxygenators. Future larger multicenter studies are required to fully discern the differences between flow-varying polymethylpentene oxygenators and their appropriate cutoffs for oxygenator dysfunction.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiovascular Institute, Allegheny Health Network, Pittsburgh, PA, USA.
- Perfusion Services, Procirca, University of Pittsburgh Medical Center, Pittsburgh, PA, USA.
- Heart and Vascular Institute, West Virginia University, Morgantown, WV, USA.
- Department of Cardiothoracic Surgery, University of Pittsburgh, Pittsburgh, PA, USA.
- Department of Critical Care Medicine, University of Pittsburgh, Pittsburgh, PA, USA.
ORCID-Profile: Shan P Modi
Zitieren
APA
Modi, S. P., D'Aloiso, B., Palmer, A., Smith, S., Arlia, P., Anselmi, M., Sanchez, P., & Ramanan, R. (2025). Comparative analysis of oxygenator dysfunction in polymethylpentene oxygenators: A pilot study. Perfusion. https://doi.org/10.1177/02676591241268402
Vancouver
Modi SP, D'Aloiso B, Palmer A, Smith S, Arlia P, Anselmi M, et al. Comparative analysis of oxygenator dysfunction in polymethylpentene oxygenators: A pilot study. Perfusion. 2025. doi:10.1177/02676591241268402
BibTeX
@article{Modi2025,
author = {Modi, Shan P and D'Aloiso, Brandon and Palmer, Amber and Smith, Stephanie and Arlia, Peter and Anselmi, Michael and Sanchez, Pablo and Ramanan, Raj},
title = {Comparative analysis of oxygenator dysfunction in polymethylpentene oxygenators: A pilot study},
journal = {Perfusion},
year = {2025},
doi = {10.1177/02676591241268402}
}
EUVIMED