Evaluation of extra‐corporeal membrane oxygenator cannulae in pulsatile and non‐pulsatile pediatric mock circuits
Artificial Organs · 2025-03 · 49 (3) · 420-430
DOI: 10.1111/aor.14897 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Artificial Organs, online erschienen am 28.10.2024. Verfasst von Lorenzo Ferrari und 5 weiteren Autor:innen. Beteiligte Einrichtung: ARTORG Center for Biomedical Engineering Research, University of Bern, Bern, Switzerland. Hauptthemen laut MEDLINE: Extracorporeal Membrane Oxygenation, Pulsatile Flow, Hemodynamics. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
This study evaluated the hemodynamic performance of arterial and venous cannulae in a compliant pediatric extracorporeal membrane oxygenation (ECMO) mock circuit in pulsatile and non‐pulsatile flow conditions.
Methods
The ECMO setup consisted of an oxygenator, diagonal pump, and standardized‐length arterial/venous tubing with pressure transducers. A validated left‐heart mock loop was adapted to simulate pediatric conditions. The pulsatile flow was driven by a computer‐controlled piston pump set at 120 bpm. A roller pump was used for non‐pulsatile conditions. The circuit was primed with 40% glycerol‐based solution. The cardiac output was set to 1 L/min and the aortic pressure to 40–50 mmHg. Four arterial cannulae (8Fr, 10Fr, 12Fr, 14Fr) and five venous cannulae (12Fr, 14Fr, 16Fr, 18Fr, 20Fr) (Medtronic, Inc., Minneapolis, MN, USA) were tested at increasing flow rate in 12 combinations.
Results
The pulsatile condition required lower ECMO pump speeds for all cannulae combinations at a given flow rate, inducing a significantly smaller increase of flow in the mock loop. Under non‐pulsatile conditions, the aortic and arterial pressures in the cannulae were higher ( p < 0.01) while no significant differences in pressure drop and pressure‐flow characteristics (M‐number) were observed. The total hemodynamic energy was higher in case of non‐pulsatile flow ( p < 0.01).
Conclusion
Under non‐pulsatile conditions, the system was characterized by overall higher pressures, resulting in higher support to the patient. The consequent increase of potential energy compensates for increases of kinetic energy, leading to a higher total hemodynamic energy. Pressure gradients and M number are independent of the testing conditions. Pulsatile testing conditions led to more physiological testing conditions, and it is recommended for ECMO testing.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- ARTORG Center for Biomedical Engineering Research, University of Bern, Bern, Switzerland.
- Department of Cardiac Surgery, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
ORCID-Profile: Lorenzo Ferrari
Zitieren
APA
Ferrari, L., Bartkevics, M., Jenni, H., Kadner, A., Siepe, M., & Obrist, D. (2025). Evaluation of extra‐corporeal membrane oxygenator cannulae in pulsatile and non‐pulsatile pediatric mock circuits. Artificial Organs, 49(3), 420-430. https://doi.org/10.1111/aor.14897
Vancouver
Ferrari L, Bartkevics M, Jenni H, Kadner A, Siepe M, Obrist D. Evaluation of extra‐corporeal membrane oxygenator cannulae in pulsatile and non‐pulsatile pediatric mock circuits. Artificial Organs. 2025;49(3):420-430. doi:10.1111/aor.14897
BibTeX
@article{Ferrari2025,
author = {Ferrari, Lorenzo and Bartkevics, Maris and Jenni, Hansjörg and Kadner, Alexander and Siepe, Matthias and Obrist, Dominik},
title = {Evaluation of extra‐corporeal membrane oxygenator cannulae in pulsatile and non‐pulsatile pediatric mock circuits},
journal = {Artificial Organs},
year = {2025},
volume = {49},
number = {3},
pages = {420--430},
doi = {10.1111/aor.14897}
}
EUVIMED