Evaluation of sample integrity and turnaround time upon implementation of a high-speed carrier-free pneumatic tube system

Hyun-Woo Lee iD, Sang-Mi Kim, Hyun-Young Kim, Soo-Youn Lee, Hyung-Doo Park iD

Annals of Clinical Biochemistry: International Journal of Laboratory Medicine · 2026

DOI: 10.1177/00045632261428924 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Annals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2026. Verfasst von Hyun-Woo Lee und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Blood Specimen Collection, Specimen Handling. Die Arbeit nennt 50 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Pneumatic tube systems (PTSs) have revolutionized sample transport efficiency, yet concerns about sample integrity remain unresolved. This study evaluated the AirLab system (PazKorea, Hanam-si, Korea), a novel high-speed carrier-free PTS, for its impact on sample integrity and turnaround time (TAT) under controlled pre-analytical conditions.

Methods

Blood specimens were collected in duplicate from 42 participants (30 healthy volunteers and 12 outpatients with chronic alcohol use) and randomly assigned to PTS or manual transport. Multiple laboratory parameters were analysed, including chemistry, hematology, coagulation, and heavy metal tests. Sample integrity was assessed by comparing test results between transport methods using both statistical analysis and clinical significance criteria based on total allowable error (TAE). Pre-analytical, analytical and total TATs were compared between transport methods.

Results

Among 20 measurands, four showed statistically significant differences: glucose, potassium (K), lactate dehydrogenase (LD), and mean corpuscular volume (MCV). PTS yielded higher results for glucose, K and LD (average percentage biases: 1.88%, 2.01%, and 15.80%, respectively), while producing lower results for MCV (average percentage bias: −0.47%). Only LD showed clinically significant differences, as its average percentage bias exceeded its desirable TAE. PTS significantly reduced both pre-analytical and total TATs for all test types, with reductions ranging from 12.12% to 22.82% and 7.32% to 8.98%, respectively, whereas analytical TATs remained unchanged.

Conclusions

The AirLab system effectively preserves sample integrity for most laboratory parameters while improving operational efficiency through TAT reduction. However, the clinically significant elevation in LD levels necessitates careful interpretation when employing this system in clinical laboratories.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood Specimen CollectionSpecimen HandlingHumansTime FactorsFemaleMaleAdultPotassiumL-Lactate Dehydrogenase

Stichwörter der Autor:innen

AirLab systemPneumatic tube systemsample integrityturnaround time

Substanzen

PotassiumL-Lactate Dehydrogenase

Autor:innen und Institutionen

  • Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Department of Laboratory Medicine, Chosun University Hospital, Chosun University School of Medicine, Gwangju, Republic of Korea.
  • Department of Medical Device Management and Research, SAIHST, Sungkyunkwan University, Seoul, Republic of Korea.

ORCID-Profile: Hyun-Woo Lee, Hyung-Doo Park

Zitieren

APA

Lee, H. W., Kim, S. M., Kim, H. Y., Lee, S. Y., & Park, H. D. (2026). Evaluation of sample integrity and turnaround time upon implementation of a high-speed carrier-free pneumatic tube system. Annals of Clinical Biochemistry: International Journal of Laboratory Medicine. https://doi.org/10.1177/00045632261428924

Vancouver

Lee HW, Kim SM, Kim HY, Lee SY, Park HD. Evaluation of sample integrity and turnaround time upon implementation of a high-speed carrier-free pneumatic tube system. Annals of Clinical Biochemistry: International Journal of Laboratory Medicine. 2026. doi:10.1177/00045632261428924

BibTeX
@article{Lee2026,
  author = {Lee, Hyun-Woo and Kim, Sang-Mi and Kim, Hyun-Young and Lee, Soo-Youn and Park, Hyung-Doo},
  title = {Evaluation of sample integrity and turnaround time upon implementation of a high-speed carrier-free pneumatic tube system},
  journal = {Annals of Clinical Biochemistry: International Journal of Laboratory Medicine},
  year = {2026},
  doi = {10.1177/00045632261428924}
}
Kostenlos starten →