EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Stability of arterial blood gas samples after delayed analysis and mechanical stress

Max Gutermuth iD, Harald Ihmsen, Frederick Krischke, Andreas Moritz, Johannes Prottengeier

PLOS One · 2025

DOI: 10.1371/journal.pone.0334710 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS One, 2025. Verfasst von Max Gutermuth und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesiology, Erlangen University Hospital, Erlangen, Germany. Hauptthemen laut MEDLINE: Blood Gas Analysis, Stress, Mechanical, Blood Specimen Collection. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To investigate the effect of prolonged time before analysis and mechanical manipulation on pre-analytical stability of biomarkers and the validity of blood gas analysis results.

Methods

We collected blood samples from 240 ICU patients from May 18, 2022 to March 31, 2023. Samples were analyzed immediately per standard operating procedure, then the syringes were kept at room temperature for 60 min, subjected to standardized mechanical forces (repeated drops) and analyzed again. Thirteen typical blood gas analytes were measured. Bland–Altman plots were prepared to assess differences between initial and delayed analyses. Differences were compared against official accuracy limits specified in German quality assurance guidelines (Rili-BAEK).

Results

For hemoglobin, creatinine, glucose, and electrolytes (calcium, sodium, chloride, bicarbonate), agreement between immediate and post-delay analyses remained within the official acceptable ranges. For pH and potassium, deviations exceeded the Rili-BAEK accuracy limits but remained clinically acceptable. Only oxygen partial pressure and lactate levels changed so markedly that they would no longer be reliable for clinical interpretation.

Conclusion

Even after a 60-min delay and excessive mechanical stress, selected blood gas analytes such as hemoglobin, glucose, and electrolytes can be considered valid. Potassium and carbon dioxide partial pressure were altered but might be suitable for approximation purposes. Findings for oxygen partial pressure and lactate were generally invalid. In the future, these findings can aid in reducing unnecessary blood sampling. These findings may guide clinicians in deciding whether repeat sampling is necessary, potentially reducing unnecessary blood draws, while reinforcing that critical parameter (pO₂, pCO₂, pH) still require prompt analysis.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood Gas AnalysisStress, MechanicalBlood Specimen CollectionHumansMaleFemaleMiddle AgedTime FactorsElectrolytesPotassiumAgedOxygenBiomarkersBlood GlucoseIntensive Care UnitsHydrogen-Ion Concentration

Substanzen

ElectrolytesPotassiumOxygenBiomarkersBlood Glucose

Autor:innen und Institutionen

  • Department of Anesthesiology, Erlangen University Hospital, Erlangen, Germany.
  • Faculty of Medicine, Friedrich Alexander University Erlangen Nuremberg, Erlangen, Germany.
  • Outpatient Surgery Center Bernard and Colleagues, Erlangen, Germany.
  • Department of Anesthesiology and Critical Care, Klagenfurt Hospital, Klagenfurt, Austria.

ORCID-Profile: Max Gutermuth

Zitieren

APA

Gutermuth, M., Ihmsen, H., Krischke, F., Moritz, A., & Prottengeier, J. (2025). Stability of arterial blood gas samples after delayed analysis and mechanical stress. PLOS One. https://doi.org/10.1371/journal.pone.0334710

Vancouver

Gutermuth M, Ihmsen H, Krischke F, Moritz A, Prottengeier J. Stability of arterial blood gas samples after delayed analysis and mechanical stress. PLOS One. 2025. doi:10.1371/journal.pone.0334710

BibTeX
@article{Gutermuth2025,
  author = {Gutermuth, Max and Ihmsen, Harald and Krischke, Frederick and Moritz, Andreas and Prottengeier, Johannes},
  title = {Stability of arterial blood gas samples after delayed analysis and mechanical stress},
  journal = {PLOS One},
  year = {2025},
  doi = {10.1371/journal.pone.0334710}
}