Safety of Contemporary Radiation Therapy in Patients With Cardiac Implantable Electronic Devices: Implications for Clinical Practice
Journal of the American Heart Association · 02.09.2025 · 14 (17)
DOI: 10.1161/jaha.125.042937 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of the American Heart Association, erschienen am 02.09.2025. Verfasst von Kaveh Karimzad und 22 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology University of Texas MD Anderson Cancer Center Houston TX USA. Hauptthemen laut MEDLINE: Defibrillators, Implantable, Pacemaker, Artificial, Radiotherapy, Neoplasms. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
Radiation therapy (RT) can cause cardiac implantable electronic device (CIED) malfunction, primarily reset. Given changes in RT and CIED technologies, large observational studies examining malfunction of contemporary CIEDs during modern‐day RT are needed to guide clinical practice.
Methods
Electronic medical records of all consecutive patients with CIEDs who underwent RT at a large tertiary cancer center between January 2015 and January 2022 were reviewed. All patients had CIED interrogation before and after completion of RT. Device‐related data and RT‐specific variables were collected. Statistical analyses were performed at the RT course level.
Results
Over a period of 7 years, 677 patients (median age, 74 [interquartile range, 67. 8–79.2] years; 72% men) with CIEDs (498 [66%] pacemakers and 254 [34%] implantable cardioverter‐defibrillators) underwent 752 courses of RT (photon, 83.5%; proton, 10.5%; electron, 6%). Only 9 patients (1.3%) underwent CIED relocation before RT. Device malfunction was observed during 8 RT courses (1.1%), primarily soft reset with data loss. All device malfunctions were observed with neutron‐producing radiation (proton and photon >10 MV). CIED type (pacemaker or implantable cardioverter‐defibrillator), magnetic resonance imaging conditional or nonconditional system, and radiation site (thorax or nonthorax) did not show statistical differences between RT courses with reset versus no reset. No transient signal interference or oversensing due to RT was recognized by the CIED.
Conclusions
CIED malfunction is a rare complication of contemporary RT, occurring primarily with neutron‐producing radiation, most of CIED malfunction cases are without serious clinical consequences. Transient signal interference and oversensing during RT does not appear to be a clinical concern in contemporary CIEDs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology University of Texas MD Anderson Cancer Center Houston TX USA.
- Division of Radiation Oncology University of Texas MD Anderson Cancer Center Houston TX USA.
- Division of Cardiology University of Texas Health Science Center at Houston, McGovern Medical School Houston TX USA.
- University of Texas Health Science Center at Houston, McGovern Medical School Houston TX USA.
- Department of Biostatistics University of Texas MD Anderson Cancer Center Houston TX USA.
- Department of Anesthesiology University of Texas MD Anderson Cancer Center Houston TX USA.
ORCID-Profile: Kaveh Karimzad, Shalin Shah, Christopher Peeler, Amit Koduri, Martin Chacon Portillo, January Tsai, Shital Vachhani, Cezar Iliescu, Bharat Kantharia, Anita Deswal
Zitieren
APA
Karimzad, K., Shah, S., Kry, S. F., Peeler, C., Pollard-Larkin, J., Koduri, A., Chacon Portillo, M., Chauhan, S., Amatullah, A., Miller, C., Strang, A., Song, J., Mir Zaman Shah, K., Lopez, D., Wagner, D., Tsai, J., Vachhani, S., Thakar, D., Iliescu, C., …, & Deswal, A. (2025). Safety of Contemporary Radiation Therapy in Patients With Cardiac Implantable Electronic Devices: Implications for Clinical Practice. Journal of the American Heart Association, 14(17). https://doi.org/10.1161/jaha.125.042937
Vancouver
Karimzad K, Shah S, Kry SF, Peeler C, Pollard-Larkin J, Koduri A, et al. Safety of Contemporary Radiation Therapy in Patients With Cardiac Implantable Electronic Devices: Implications for Clinical Practice. Journal of the American Heart Association. 2025;14(17). doi:10.1161/jaha.125.042937
BibTeX
@article{Karimzad2025,
author = {Karimzad, Kaveh and Shah, Shalin and Kry, Stephen F and Peeler, Christopher and Pollard-Larkin, Julianne and Koduri, Amit and Chacon Portillo, Martin and Chauhan, Siddharth and Amatullah, Atia and Miller, Chase and Strang, Amanda and Song, Juhee and Mir Zaman Shah, Khatera and Lopez, Donelo and Wagner, Darla and Tsai, January and Vachhani, Shital and Thakar, Dilip and Iliescu, Cezar and Durand, Jean-Bernard and Dougherty, Anne H and Kantharia, Bharat and Deswal, Anita},
title = {Safety of Contemporary Radiation Therapy in Patients With Cardiac Implantable Electronic Devices: Implications for Clinical Practice},
journal = {Journal of the American Heart Association},
year = {2025},
volume = {14},
number = {17},
doi = {10.1161/jaha.125.042937}
}
EUVIMED