Periprocedural Assessment of Paravalvular Regurgitation After Transcatheter Aortic Valve Replacement Using Diastolic Delta and Videodensitometry
Journal of the American Heart Association · 17.09.2024 · 13 (18)
DOI: 10.1161/jaha.124.035587 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of the American Heart Association, erschienen am 17.09.2024. Verfasst von Maxim J P Rooijakkers und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology Radboud University Medical Center Nijmegen The Netherlands. Hauptthemen laut MEDLINE: Aortic Valve Insufficiency, Transcatheter Aortic Valve Replacement. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
Paravalvular regurgitation (PVR) is a common complication after transcatheter aortic valve replacement, posing an increased risk of heart failure and mortality. Accurate intraprocedural quantification of PVR is challenging. Both hemodynamic indices and videodensitometry can be used for intraprocedural assessment of PVR. We compared the predictive value of the isolated versus combined use of the hemodynamic index diastolic delta (DD) and videodensitometry for the incidence of relevant PVR 1 month after transcatheter aortic valve replacement. Methods and Results In this prospective cohort study, patients underwent periprocedural PVR assessment by DD and videodensitometry (using left ventricular outflow tract—aortic regurgitation [LVOT‐AR]). Cardiac magnetic resonance served as reference modality for PVR assessment. Relevant PVR was defined as cardiac magnetic resonance–regurgitant fraction >20%. Fifty‐one patients were enrolled in this study. Mean age was 80.6±5.2 years and 45.1% of patients were men. Mean LVOT‐AR and cardiac magnetic resonance–regurgitant fraction were 8.2%±7.8% and 11.7%±9.6%, respectively. The correlation between DD and LVOT‐AR was weak ( r =−0.36). DD and LVOT‐AR showed a comparable accuracy to predict relevant PVR (area under the curve 0.82, 95% CI: 0.69–0.95 versus area area under the time–density curve 0.80, 95% CI: 0.62–0.99). The combination of DD and LVOT‐AR improved the prediction of relevant PVR (area under the time–density curve, 0.90, 95% CI: 0.81–0.99), and resulted in an increased concordance (86.3%) and positive predictive value (75%) compared with DD alone (76.5% and 40%, respectively), or LVOT‐AR alone (82.3% and 50%, respectively).
Conclusions
DD and videodensitometry are both accurate and feasible modalities for the assessment of PVR after transcatheter aortic valve replacement. The synergistic use of both techniques increases the predictive value for relevant PVR after transcatheter aortic valve replacement. Registration URL: https://www.clinicaltrials.gov ; Unique identifier: NCT04281771.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology Radboud University Medical Center Nijmegen The Netherlands.
- CORRIB Research Centre for Advanced Imaging and Core Laboratory University of Galway Galway Ireland.
- Islamic Center of Cardiology and Cardiac Surgery Al-Azhar University Cairo Egypt.
- Department of Medical BioSciences Radboud University Medical Center Nijmegen The Netherlands.
- Department of Health Sciences, section Biostatistics Radboud Institute for Health Sciences Nijmegen The Netherlands.
- Department of Cardiothoracic Surgery Radboud University Medical Center Nijmegen The Netherlands.
ORCID-Profile: Maxim J P Rooijakkers, Ahmed Elkoumy, Niels A Stens, Marleen H van Wely, Geert A A Versteeg, Tsung-Ying Tsai, Robin H Heijmen, Patrick W Serruys, Osama Soliman, Niels van Royen
Zitieren
APA
Rooijakkers, M. J. P., Elkoumy, A., Stens, N. A., van Wely, M. H., Versteeg, G. A. A., Tsai, T. Y., Rodwell, L., Heijmen, R. H., Serruys, P. W., Soliman, O., & van Royen, N. (2024). Periprocedural Assessment of Paravalvular Regurgitation After Transcatheter Aortic Valve Replacement Using Diastolic Delta and Videodensitometry. Journal of the American Heart Association, 13(18). https://doi.org/10.1161/jaha.124.035587
Vancouver
Rooijakkers MJP, Elkoumy A, Stens NA, van Wely MH, Versteeg GAA, Tsai TY, et al. Periprocedural Assessment of Paravalvular Regurgitation After Transcatheter Aortic Valve Replacement Using Diastolic Delta and Videodensitometry. Journal of the American Heart Association. 2024;13(18). doi:10.1161/jaha.124.035587
BibTeX
@article{Rooijakkers2024,
author = {Rooijakkers, Maxim J P and Elkoumy, Ahmed and Stens, Niels A and van Wely, Marleen H and Versteeg, Geert A A and Tsai, Tsung-Ying and Rodwell, Laura and Heijmen, Robin H and Serruys, Patrick W and Soliman, Osama and van Royen, Niels},
title = {Periprocedural Assessment of Paravalvular Regurgitation After Transcatheter Aortic Valve Replacement Using Diastolic Delta and Videodensitometry},
journal = {Journal of the American Heart Association},
year = {2024},
volume = {13},
number = {18},
doi = {10.1161/jaha.124.035587}
}
EUVIMED