EUVIMEDEUVIMED entdecken
Crossref

Commissural Alignment in SAPIEN 3 Valves: Impact on Gradient and Mortality at Follow‐Up

Leire Unzue iD, Belén Díaz Antón, Adrián Peláez Laderas, Lorena Martín Polo, Jose María Castellano Vázquez, Francisco José Rodríguez Rodrigo, Francisco Javier Parra Jimenez, Leticia Fernández-Friera, Rodrigo Teijeiro Mestre, Eulogio García Fernández

Catheterization and Cardiovascular Interventions · 2025

DOI: 10.1002/ccd.31418 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Catheterization and Cardiovascular Interventions, 2025. Verfasst von Leire Unzue und 9 weiteren Autor:innen. Beteiligte Einrichtung: Departamento de Cardiología Intervencionista, HM CIEC MADRID (Centro Integral de Enfermedades Cardiovasculares), HM Hospitales, Madrid, Spain. Hauptthemen laut MEDLINE: Heart Valve Prosthesis, Aortic Valve, Aortic Valve Stenosis, Hemodynamics. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Commissural misalignment (CMA) has demonstrated to difficult future coronary access in self‐expandable prosthesis and has been related with subclinical leaflet thrombosis and a potential increase in valve gradients and aortic regurgitation at follow‐up.

Aims

The aim of our study was to assess the commissural alignment of the balloon expandable SAPIEN 3 valve (Edwards Lifesciences) and to evaluate its impact on transvalvular gradient and mortality at 5 years of follow‐up.

Methods

Consecutive patients treated with SAPIEN 3 prosthesis guided with intraprocedural transesophageal echocardiogram (TEE) with a minimal follow‐up of 5 years were included. Commissural alignment between the prosthesis and the native valve was measured by TEE by two independent cardiologists using a standardized 3‐cusp view pre and post‐valve deployment. CMA was defined as a neocommissure deviation > 30° compared with native commissures. Echocardiographic gradients and clinical events were recorded during follow‐up and compared between groups.

Results

One hundred and five patients (54% male, mean age 80.4 years) were included; 63 (40%) were classified as “aligned” (CMA < 30°) and 42 (60%) as “misaligned” (CMA > 30°). Patients with CMA > 30° presented smaller aortic annulus (23.8 vs. 20.7, p < 0.01) and received smaller valves. There was a slight increase of mean transvalvular gradient during follow‐up (9.2−10.3 mmHg, p < 0.05), without significant differences between groups. Mortality rate was 16.2% at 1 year and 49.5% at 5 years; and not related to commissural alignment.

Conclusions

Significant CMA is present in 40% of patients treated with SAPIEN 3 valves, however, we didn't find a significant association between CMA and gradient increase or mortality during long‐term follow‐up.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart Valve ProsthesisAortic ValveAortic Valve StenosisHemodynamicsTranscatheter Aortic Valve ReplacementHeart Valve Prosthesis ImplantationHumansMaleFemaleProsthesis DesignTreatment OutcomeTime FactorsEchocardiography, TransesophagealAgedAged, 80 and overRisk FactorsAortic Valve InsufficiencyRetrospective StudiesUltrasonography, Interventional

Stichwörter der Autor:innen

SAPIEN 3 valvesTAVRcommissural alignment

Autor:innen und Institutionen

  • Departamento de Cardiología Intervencionista, HM CIEC MADRID (Centro Integral de Enfermedades Cardiovasculares), HM Hospitales, Madrid, Spain.
  • Instituto de Investigación Sanitaria, HM Hospitales, Madrid, Spain.
  • Departamento de imagen cardiaca, HM CIEC MADRID (Centro Integral de Enfermedades Cardiovasculares), HM Hospitales, Madrid, Spain.
  • Facultad HM de Ciencias de la Salud de la Universidad Camilo José Cela, Madrid, Spain.
  • Centro de Investigación en Red de Enfermedades Respiratorias (CIBERES), Instituto de Salud Carlos III (ISCIII), Madrid, Spain.
  • Departamento de Cardiología, HM CIEC MADRID (Centro Integral de Enfermedades Cardiovasculares), HM Hospitales, Madrid, Spain.

ORCID-Profile: Leire Unzue

Zitieren

APA

Unzue, L., Antón, B. D., Laderas, A. P., Polo, L. M., Vázquez, J. M. C., Rodrigo, F. J. R., Jimenez, F. J. P., Fernández-Friera, L., Mestre, R. T., & Fernández, E. G. (2025). Commissural Alignment in SAPIEN 3 Valves: Impact on Gradient and Mortality at Follow‐Up. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31418

Vancouver

Unzue L, Antón BD, Laderas AP, Polo LM, Vázquez JMC, Rodrigo FJR, et al. Commissural Alignment in SAPIEN 3 Valves: Impact on Gradient and Mortality at Follow‐Up. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31418

BibTeX
@article{Unzue2025,
  author = {Unzue, Leire and Antón, Belén Díaz and Laderas, Adrián Peláez and Polo, Lorena Martín and Vázquez, Jose María Castellano and Rodrigo, Francisco José Rodríguez and Jimenez, Francisco Javier Parra and Fernández-Friera, Leticia and Mestre, Rodrigo Teijeiro and Fernández, Eulogio García},
  title = {Commissural Alignment in SAPIEN 3 Valves: Impact on Gradient and Mortality at Follow‐Up},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2025},
  doi = {10.1002/ccd.31418}
}
Kostenlos starten →