Effect of Sequential, Colocalized Radiofrequency and Pulsed Field Ablation on Cardiac Lesion Size and Histology

Atul Verma iD, Jennifer Maffre iD, Tushar Sharma, Salman Farshchi-Heydari iD

Circulation: Arrhythmia and Electrophysiology · 2025-02 · 18 (2)

DOI: 10.1161/circep.124.013143 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Circulation: Arrhythmia and Electrophysiology, 2025-02. Verfasst von Atul Verma und 3 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiology, McGill University Health Centre, Montreal, QC, Canada (A.V.). Hauptthemen laut MEDLINE: Catheter Ablation, Myocardium, Heart Ventricles, Pulsed Radiofrequency Treatment. Die Arbeit wurde bisher 57-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Sequential application of radiofrequency with pulsed field (PF) ablation may increase lesion depth while preserving the advantages of PF. The study’s aim was to determine lesion dimensions of sequential, colocalized radiofrequency and PF ablation. METHODS: A preclinical study using swine (n=4) performed lesions in the right/left ventricles. Ablations were performed with a force-sensing 3.5-mm irrigated-tip ablation catheter using a generator delivering both radiofrequency and PF. PF was delivered using unipolar, biphasic pulses at a standard dose (PF index, 300) with 4-mL/min irrigation. Radiofrequency was delivered at 50 W for 10 s (15 mL/min). Lesions were created by applying colocalized radiofrequency followed by sequential application of PF on the same location, PF followed by sequential application of radiofrequency on the same location, PF alone, or radiofrequency alone. Tissue was collected after 2 hours for lesion assessment. Results are mean±SD. RESULTS: Forty-five lesions were analyzed. The lesion depth of radiofrequency alone was 4.9±0.8 mm. The mean lesion depth and width for PF alone were 3.5±0.6 and 5.1±1.8 mm. Lesion depths for combined applications were significantly greater versus PF alone (6.2±1.8 mm radiofrequency followed by sequential application of PF on the same location; 5.7±1.3 mm PF followed by sequential application of radiofrequency on the same location; P

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Catheter AblationMyocardiumHeart VentriclesPulsed Radiofrequency TreatmentAnimalsSwineTime FactorsModels, Animal

Stichwörter der Autor:innen

ablationcathetersheart atriaheart ventriclesirreversible electroporation therapymodels, animaltechnology

Autor:innen und Institutionen

  • Division of Cardiology, McGill University Health Centre, Montreal, QC, Canada (A.V.).
  • Department of Research and Development, Biosense Webster, Inc, Irvine, CA (J.M., T.S., S.F.-H.).

ORCID-Profile: Atul Verma, Jennifer Maffre, Salman Farshchi-Heydari

Zitieren

APA

Verma, A., Maffre, J., Sharma, T., & Farshchi-Heydari, S. (2025). Effect of Sequential, Colocalized Radiofrequency and Pulsed Field Ablation on Cardiac Lesion Size and Histology. Circulation: Arrhythmia and Electrophysiology, 18(2). https://doi.org/10.1161/circep.124.013143

Vancouver

Verma A, Maffre J, Sharma T, Farshchi-Heydari S. Effect of Sequential, Colocalized Radiofrequency and Pulsed Field Ablation on Cardiac Lesion Size and Histology. Circulation: Arrhythmia and Electrophysiology. 2025;18(2). doi:10.1161/circep.124.013143

BibTeX
@article{Verma2025,
  author = {Verma, Atul and Maffre, Jennifer and Sharma, Tushar and Farshchi-Heydari, Salman},
  title = {Effect of Sequential, Colocalized Radiofrequency and Pulsed Field Ablation on Cardiac Lesion Size and Histology},
  journal = {Circulation: Arrhythmia and Electrophysiology},
  year = {2025},
  volume = {18},
  number = {2},
  doi = {10.1161/circep.124.013143}
}
Kostenlos starten →