Effect of Implant Length and Insertion Depth on Primary Stability of Short Dental Implant

Hsiang-Chun Wu, Heng-Li Huang, Lih-Jyh Fuh, Ming-Tzu Tsai, Jui-Ting Hsu

International Journal of Oral and Maxillofacial Implants · 2023-01 · 38 (1) · 62-70

DOI: 10.11607/jomi.9769 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Oral and Maxillofacial Implants, 2023-01. Verfasst von Hsiang-Chun Wu und 4 weiteren Autor:innen. Hauptthemen laut MEDLINE: Dental Implants. Die Arbeit wurde bisher 14-mal zitiert (Crossref).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To evaluate the effect of insertion depth, bone type, and implant diameter on the primary stability of short implants. Materials and Methods: Commercial dental implants with different lengths (6 and 8 mm; BLX, Straumann) were inserted into artificial bone specimens of good and poor quality at three different depth positions: equicrestal, 1-mm subcrestal, and 2-mm subcrestal. Insertion torque values were recorded spontaneously during the implant procedure. Both maximum insertion torque values (MITVs) and final insertion torque values (FITVs) were recorded. Subsequently, Periotest values (PTVs) and implant stability quotients (ISQs) were measured for all specimens.

Results

The mean MITVs of all groups ranged from 31.8 to 46.2 Ncm. However, the mean FITVs of all groups ranged from 8.8 to 29 Ncm. Torque values decreased significantly when the implants were inserted into their final positions. When insertion depth was increased, the PTV and ISQ decreased. Long implants and implants inserted into good-quality bone yielded greater primary stability, and bone quality appeared to have a greater effect on primary stability.

Conclusion

When 6-mm short implants are inserted in a subcrestal position, low primary stability may be yielded, particularly in poor-quality bone.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Dental ImplantsDental Implantation, EndosseousTorqueDental Prosthesis Design

Substanzen

Dental Implants

Zitieren

APA

Wu, H. C., Huang, H. L., Fuh, L. J., Tsai, M. T., & Hsu, J. T. (2023). Effect of Implant Length and Insertion Depth on Primary Stability of Short Dental Implant. International Journal of Oral and Maxillofacial Implants, 38(1), 62-70. https://doi.org/10.11607/jomi.9769

Vancouver

Wu HC, Huang HL, Fuh LJ, Tsai MT, Hsu JT. Effect of Implant Length and Insertion Depth on Primary Stability of Short Dental Implant. International Journal of Oral and Maxillofacial Implants. 2023;38(1):62-70. doi:10.11607/jomi.9769

BibTeX
@article{Wu2023,
  author = {Wu, Hsiang-Chun and Huang, Heng-Li and Fuh, Lih-Jyh and Tsai, Ming-Tzu and Hsu, Jui-Ting},
  title = {Effect of Implant Length and Insertion Depth on Primary Stability of Short Dental Implant},
  journal = {International Journal of Oral and Maxillofacial Implants},
  year = {2023},
  volume = {38},
  number = {1},
  pages = {62--70},
  doi = {10.11607/jomi.9769}
}
Kostenlos starten →