Effect of Implant Length and Insertion Depth on Primary Stability of Short Dental Implant
International Journal of Oral and Maxillofacial Implants · 2023-01 · 38 (1) · 62-70
DOI: 10.11607/jomi.9769 ↗At a glance
Journal article in International Journal of Oral and Maxillofacial Implants, 2023-01. Written by Hsiang-Chun Wu and 4 co-authors. Major MEDLINE topics: Dental Implants. Cited 14 times so far (Crossref).
Abstract (original language)
Purpose
To evaluate the effect of insertion depth, bone type, and implant diameter on the primary stability of short implants. Materials and Methods: Commercial dental implants with different lengths (6 and 8 mm; BLX, Straumann) were inserted into artificial bone specimens of good and poor quality at three different depth positions: equicrestal, 1-mm subcrestal, and 2-mm subcrestal. Insertion torque values were recorded spontaneously during the implant procedure. Both maximum insertion torque values (MITVs) and final insertion torque values (FITVs) were recorded. Subsequently, Periotest values (PTVs) and implant stability quotients (ISQs) were measured for all specimens.
Results
The mean MITVs of all groups ranged from 31.8 to 46.2 Ncm. However, the mean FITVs of all groups ranged from 8.8 to 29 Ncm. Torque values decreased significantly when the implants were inserted into their final positions. When insertion depth was increased, the PTV and ISQ decreased. Long implants and implants inserted into good-quality bone yielded greater primary stability, and bone quality appeared to have a greater effect on primary stability.
Conclusion
When 6-mm short implants are inserted in a subcrestal position, low primary stability may be yielded, particularly in poor-quality bone.
Subjects
MeSH terms (MEDLINE)
Substances
Cite
APA
Wu, H. C., Huang, H. L., Fuh, L. J., Tsai, M. T., & Hsu, J. T. (2023). Effect of Implant Length and Insertion Depth on Primary Stability of Short Dental Implant. International Journal of Oral and Maxillofacial Implants, 38(1), 62-70. https://doi.org/10.11607/jomi.9769
Vancouver
Wu HC, Huang HL, Fuh LJ, Tsai MT, Hsu JT. Effect of Implant Length and Insertion Depth on Primary Stability of Short Dental Implant. International Journal of Oral and Maxillofacial Implants. 2023;38(1):62-70. doi:10.11607/jomi.9769
BibTeX
@article{Wu2023,
author = {Wu, Hsiang-Chun and Huang, Heng-Li and Fuh, Lih-Jyh and Tsai, Ming-Tzu and Hsu, Jui-Ting},
title = {Effect of Implant Length and Insertion Depth on Primary Stability of Short Dental Implant},
journal = {International Journal of Oral and Maxillofacial Implants},
year = {2023},
volume = {38},
number = {1},
pages = {62--70},
doi = {10.11607/jomi.9769}
}
EUVIMED