Mechanical Influence of Edentulous Mandibular Morphology on Peri-implant Bone in Implant Prosthetics: Threedimensional Finite Element Analysis
The International Journal of Oral & Maxillofacial Implants · 2024
DOI: 10.11607/jomi.10859 ↗At a glance
Journal article in The International Journal of Oral & Maxillofacial Implants, 2024. Written by Takumi Hamada and 4 co-authors. Major MEDLINE topics: Finite Element Analysis, Jaw, Edentulous, Mandible, Imaging, Three-Dimensional.
Abstract (original language)
PURPOSE: To examine the mechanical influence of edentulous mandible morphology on peri-implant bone in implant prosthetics via finite element analysis (FEA). MATERIALS AND METHODS: Computed tomographic (CT) data from 25 patients with edentulous mandibles were selected. Measurements such as the radius of mandibular curvature and the height of the mandible at the midline were taken to clarify the morphologic characteristics of the mandible. From the measurements, two patients with the smallest and largest radii of the mandible were selected. Two types of 3D finite element models were created from the two selected edentulous mandibles (small and large radius), and they included superstructures (a cantilever bridge), implants, and abutments. In both models, four implants were placed between the right and left mental foramina. The upper surface of the mandibular condyle was constrained, and a vertical load of 100 N was applied on the occlusal surface of the right first molar. 3D FEA of each model was performed to examine the mechanical influence of the edentulous mandible morphology on the peri-implant bone. RESULTS: The mandibular morphology measurement in the CT images indicated that the smaller the mandibular height, the larger the radius of the anterior mandibular curvature. FEA revealed that higher equivalent stress was generated in the model of the peri-implant bone with a larger radiusof curvature than that of the model with a smaller radius of curvature. The highest equivalent stress in the mandible was generated in the distal margin of the peri-implant bone posterior to the loaded side of the large-radius model. CONCLUSIONS: The mandibular morphology had a mechanical influence on the peri-implant bone.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Cite
APA
Hamada, T., Gonda, T., Murase, K., Maeda, Y., & Ikebe, K. (2024). Mechanical Influence of Edentulous Mandibular Morphology on Peri-implant Bone in Implant Prosthetics: Threedimensional Finite Element Analysis. The International Journal of Oral & Maxillofacial Implants. https://doi.org/10.11607/jomi.10859
Vancouver
Hamada T, Gonda T, Murase K, Maeda Y, Ikebe K. Mechanical Influence of Edentulous Mandibular Morphology on Peri-implant Bone in Implant Prosthetics: Threedimensional Finite Element Analysis. The International Journal of Oral & Maxillofacial Implants. 2024. doi:10.11607/jomi.10859
BibTeX
@article{Hamada2024,
author = {Hamada, Takumi and Gonda, Tomoya and Murase, Kohei and Maeda, Yoshinobu and Ikebe, Kazunori},
title = {Mechanical Influence of Edentulous Mandibular Morphology on Peri-implant Bone in Implant Prosthetics: Threedimensional Finite Element Analysis},
journal = {The International Journal of Oral & Maxillofacial Implants},
year = {2024},
doi = {10.11607/jomi.10859}
}
EUVIMED