The Distortion of the Implant-Framework Interface of Screw- Retained Co–Cr Ceramic Fixed Dental Prosthesis Following Ceramic Veneering: An In Vitro Study
The International Journal of Prosthodontics · 2025
DOI: 10.11607/ijp.9299 ↗At a glance
Journal article in The International Journal of Prosthodontics, 2025. Written by Marco Toia and 4 co-authors. Major MEDLINE topics: Dental Prosthesis, Implant-Supported, Dental Veneers, Chromium Alloys, Denture, Partial, Fixed.
Abstract (original language)
PURPOSE: To evaluate the effect of ceramic firing cycles on the interface of screw-retained implant-supported metal-ceramic fixed dental prostheses (FDPs). MATERIALS AND METHODS: A total of 22 frameworks were produced in cobalt-chrome (CoCr) alloy using an integrated CAD/CAM system. The frameworks were manufactured with an additive process on the milled implant-framework interface (IFI). The total distortion of the interface was measured before (T1) and after (T2) the ceramic firing thermal cycles. The interface distortion was measured using a coordinate measuring machine (CMM) with an uncertainty of 2.5 nm, a resolution of 25 nm, and a universal software for dimensional metrology. To test whether there was a significant difference in total 3D distortion, Kruskal-Wallis test and Dunn test were employed; linear distortions between groups were compared using Kolmogorov-Smirnov test and Mann-Whitney test. A P value of ≤ .05 was considered statistically significant. RESULTS: The total 3D distortion between the nominal value and T1 was 0.5 ± 1.0 μm, between the nominal value and T2 was 10.5 ± 4.9 μm, and between T1 and T2 was 10.6 ± 4.9 μm. The distortion between T1 and the nominal value was significantly smaller than that between the other two groups. The average deviations around the three axes and diameters were not significantly different among the three groups. CONCLUSIONS: CAD/CAM manufacturing of the IFI is an accurate process and does not present appreciable distortions. The ceramic firing cycles induce 3D-interface distortion. The observed distortion values were well below clinically relevant levels.
Subjects
MeSH terms (MEDLINE)
Substances
Cite
APA
Toia, M., Parpaiola, A., Papia, E., Rodriguez Y Baena, R., & Lupi, S. M. (2025). The Distortion of the Implant-Framework Interface of Screw- Retained Co–Cr Ceramic Fixed Dental Prosthesis Following Ceramic Veneering: An In Vitro Study. The International Journal of Prosthodontics. https://doi.org/10.11607/ijp.9299
Vancouver
Toia M, Parpaiola A, Papia E, Rodriguez Y Baena R, Lupi SM. The Distortion of the Implant-Framework Interface of Screw- Retained Co–Cr Ceramic Fixed Dental Prosthesis Following Ceramic Veneering: An In Vitro Study. The International Journal of Prosthodontics. 2025. doi:10.11607/ijp.9299
BibTeX
@article{Toia2025,
author = {Toia, Marco and Parpaiola, Andrea and Papia, Evaggelia and Rodriguez Y Baena, Ruggero and Lupi, Saturnino Marco},
title = {The Distortion of the Implant-Framework Interface of Screw- Retained Co–Cr Ceramic Fixed Dental Prosthesis Following Ceramic Veneering: An In Vitro Study},
journal = {The International Journal of Prosthodontics},
year = {2025},
doi = {10.11607/ijp.9299}
}
EUVIMED