Comparative in vitro evaluation of microgap in titanium stock versus cobalt–chrome custom abutments on a conical connection implant: Effect of crown cementation and ceramic veneering
Clinical Oral Implants Research · 2024
DOI: 10.1111/clr.14317 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Clinical Oral Implants Research, 2024. Verfasst von Federico Herrero-Climent und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Conservative and Prosthetic Dentistry, Faculty of Odontology, Complutense University of Madrid, Madrid, Spain. Hauptthemen laut MEDLINE: Titanium, Dental Veneers, Dental Abutments, Computer-Aided Design. Die Arbeit nennt 66 Literaturangaben. Gefördert von: SOADCO SL, Spain.
Abstract (Originalsprache)
Objective
To compare the implant–abutment connection microgap between computer‐aided design and computer‐aided manufacturing (CAD/CAM) milled or laser‐sintered cobalt–chrome custom abutments with or without ceramic veneering and titanium stock abutments with or without crown cementation. Material and Methods Six groups of six abutments each were prepared: (1) CAD/CAM cobalt–chrome custom abutments: milled, milled with ceramic veneering, laser‐sintered, and laser‐sintered with ceramic veneering (four groups: MIL, MIL‐C, SIN, and SIN‐C, respectively) and (2) titanium stock abutments with or without zirconia crown cementation (two groups: STK and STK‐Z, respectively). Abutments were screwed to the implants by applying 30 Ncm torque. All 36 samples were sectioned along their long axes. The implant–abutment connection microgap was measured using scanning electron microscopy on the right and left sides of the connection at the upper, middle, and lower levels. Data were analyzed using the Kruskal–Wallis test ( p < .05).
Results
Mean values (μm) of the microgap were 0.54 ± 0.44 (STK), 0.55 ± 0.48 (STK‐Z), 1.53 ± 1.30 (MIL), 2.30 ± 2.2 (MIL‐C), 1.53 ± 1.37 (SIN), and 1.87 ± 1.8 (SIN‐C). Although significant differences were observed between the STK and STK‐Z groups and the other groups ( p < .05), none were observed between the milled and laser‐sintered groups before or after ceramic veneering. The largest microgap was observed at the upper level in all groups.
Conclusions
Titanium stock abutments provided a closer fit than cobalt–chrome custom abutments. Neither crown cementation nor ceramic veneering resulted in significant changes in the implant–abutment connection microgap.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Conservative and Prosthetic Dentistry, Faculty of Odontology, Complutense University of Madrid, Madrid, Spain.
ORCID-Profile: Francisco Martínez-Rus, María Paz Salido, Guillermo Pradíes
Förderung
- SOADCO SL, Spain
Zitieren
APA
Herrero-Climent, F., Martínez-Rus, F., Salido, M. P., Roldán, D., & Pradíes, G. (2024). Comparative in vitro evaluation of microgap in titanium stock versus cobalt–chrome custom abutments on a conical connection implant: Effect of crown cementation and ceramic veneering. Clinical Oral Implants Research. https://doi.org/10.1111/clr.14317
Vancouver
Herrero-Climent F, Martínez-Rus F, Salido MP, Roldán D, Pradíes G. Comparative in vitro evaluation of microgap in titanium stock versus cobalt–chrome custom abutments on a conical connection implant: Effect of crown cementation and ceramic veneering. Clinical Oral Implants Research. 2024. doi:10.1111/clr.14317
BibTeX
@article{HerreroCliment2024,
author = {Herrero-Climent, Federico and Martínez-Rus, Francisco and Salido, María Paz and Roldán, David and Pradíes, Guillermo},
title = {Comparative in vitro evaluation of microgap in titanium stock versus cobalt–chrome custom abutments on a conical connection implant: Effect of crown cementation and ceramic veneering},
journal = {Clinical Oral Implants Research},
year = {2024},
doi = {10.1111/clr.14317}
}
EUVIMED