Cie L*A*B* Values in VITA Classical and VITA 3D Master by Two Dental Spectrophotometers

Susana Beatriz Ferreira Dias, João Miguel Lourenço Silveira, Ruben Miguel Nunes Pereira, Ana Beatriz Cardoso, António Duarte Sola Pereira da Mata, Duarte Nuno da Silva Marques

International Journal of Prosthodontics · 2023-10 · 36 (5)

DOI: 10.11607/ijp.7235 ↗

At a glance

Journal article in International Journal of Prosthodontics, 2023-10. Written by Susana Beatriz Ferreira Dias and 5 co-authors. Major MEDLINE topics: Dental Prosthesis Design, Prosthesis Coloring. Cited 5 times so far (Crossref).

SantraukaSubjectsBibliografiniai duomenysCiteRelated studies

Abstract (original language)

PURPOSE: To compare the CIE L*a*b* values of two different dental shade guides using two shade-matching spectrophotometers. MATERIALS AND METHODS: SpectroShade Micro (SS) and Easyshade (ES) (VITA Zahnfabrik) were tested. One calibrated operator performed 30 measurements of each tab from three batches of the two guides, VITA Classical (VC) and VITA 3D Shade Master (VM). Receiver operating characteristic (ROC) curves and intraclass coefficients (ICC) between the different batches were calculated. Results were given as mean and SD of the L*a*b* values and the respective color differences according to the CIEDE 2000 formula (ΔE00) for each of the shade tabs in the two spectrophotometers and then analyzed with independent student t test (α = .05). RESULTS: A total of 1,440 and 2,610 measurements for VC and VM, respectively, were performed per spectrophotometer. There were statistically significant differences between the two devices for all L*a*b* values, with the exception of: L* for 3R1.5; a* for 2R1.5, 3L1.5, 2L2.5, 3M2, 3L2.5, 4L1.5, and 4L2.5; and b* for D3, 1M2, and 3M2. When assessing the same shade guide, differences in ΔE00 were detected between devices above the acceptability threshold (AT) (ΔE00 ≥ 1.8) for all shade tabs except for VC (C1, C2, D3, A3.5, C3, and A4) and VM (4M1, 3R2.5, 4L2.5, and 4M3). The overall mean of the interdevice ΔE00 was 2.2 ± 1.0 for VC and 2.5 ± 1.0 for VM. CONCLUSIONS: The two dental spectrophotometers presented high ICC and ROC values, which validates their indication as auxiliary tools. However, there are discrepancies in the hues D (VC) and R (VM) with interdevice variability when evaluated for the L*a*b* component. Int J Prosthodont 2023;36:e38-e52.

Subjects

MeSH terms (MEDLINE)

Dental Prosthesis DesignProsthesis ColoringColorSpectrophotometryColorimetry

Cite

APA

Dias, S. B. F., Silveira, J. M. L., Pereira, R. M. N., Cardoso, A. B., Pereira da Mata, A. D. S., & Marques, D. N. D. S. (2023). Cie L*A*B* Values in VITA Classical and VITA 3D Master by Two Dental Spectrophotometers. International Journal of Prosthodontics, 36(5). https://doi.org/10.11607/ijp.7235

Vancouver

Dias SBF, Silveira JML, Pereira RMN, Cardoso AB, Pereira da Mata ADS, Marques DNDS. Cie L*A*B* Values in VITA Classical and VITA 3D Master by Two Dental Spectrophotometers. International Journal of Prosthodontics. 2023;36(5). doi:10.11607/ijp.7235

BibTeX
@article{Dias2023,
  author = {Dias, Susana Beatriz Ferreira and Silveira, João Miguel Lourenço and Pereira, Ruben Miguel Nunes and Cardoso, Ana Beatriz and Pereira da Mata, António Duarte Sola and Marques, Duarte Nuno da Silva},
  title = {Cie L*A*B* Values in VITA Classical and VITA 3D Master by Two Dental Spectrophotometers},
  journal = {International Journal of Prosthodontics},
  year = {2023},
  volume = {36},
  number = {5},
  doi = {10.11607/ijp.7235}
}
Start for free →