EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY 4.0

Clinical Accuracy of an Intraoral Scanner for Tooth Shade Selection Compared With a Spectrophotometer and Visual Matching: A Comparative Study

Amr Sameer Baqleh iD, Nabil Alhouri

Clinical and Experimental Dental Research · 2026

DOI: 10.1002/cre2.70347 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Clinical and Experimental Dental Research, 2026. Verfasst von Amr Sameer Baqleh und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Fixed Prosthodontics, Faculty of Dentistry, Damascus University, Damascus, Syria. Hauptthemen laut MEDLINE: Prosthesis Coloring, Tooth, Dental Prosthesis Design. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

This study evaluated the performance of an intraoral scanner (Medit i700) for tooth shade selection by comparing it with a spectrophotometric reference device (VITA Easyshade) and visual observers, using VITA Classical and VITA 3D‐Master shade guides. Material and Methods Sixteen patients (48 right maxillary teeth: central incisor, canine, and first molar) were recruited from the fixed prosthodontics clinic. For each tooth, shade was recorded three times with Medit i700, three visual observers, and VITA Easyshade under standardized daylight conditions using both VITA Classical and VITA 3D‐Master guides. Easyshade was used as the reference. Exact shade matching between each method and the reference was calculated as percentages. A secondary analysis focused on the value (brightness) component, assessing exact value match and value acceptability within ± 1 value level. Chi‐square tests were applied to explore the influence of method, shade guide, and tooth type on value outcomes ( α = 0.05).

Results

Exact shade matching was low for all methods and both shade guides. When only value was considered, agreement increased substantially: the intraoral scanner and visual observers achieved moderate exact value matches and high value acceptability (typically > 80% within ± 1 value level of the reference) across both shade guides. Chi‐square analysis showed no significant differences in value performance between methods; however, shade guide type showed a significant effect on exact value matching, while value acceptability did not differ. In contrast, tooth type had a statistically significant influence on value outcomes, with higher agreement in molars (and, to a lesser extent, canines) than in incisors ( p < 0.05).

Conclusions

Within the limitations of this clinical study, the intraoral scanner Medit i700 did not provide sufficient accuracy for full shade selection, but showed high agreement for value determination. Intraoral scanners should be integrated with spectrophotometric and controlled visual assessment for shade matching in esthetic restorations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Prosthesis ColoringToothDental Prosthesis DesignHumansSpectrophotometryFemaleIncisorColorMaleCuspidAdult

Stichwörter der Autor:innen

intraoral scannershade matchingspectrophotometervalue

Autor:innen und Institutionen

  • Department of Fixed Prosthodontics, Faculty of Dentistry, Damascus University, Damascus, Syria.

ORCID-Profile: Amr Sameer Baqleh

Zitieren

APA

Baqleh, A. S., & Alhouri, N. (2026). Clinical Accuracy of an Intraoral Scanner for Tooth Shade Selection Compared With a Spectrophotometer and Visual Matching: A Comparative Study. Clinical and Experimental Dental Research. https://doi.org/10.1002/cre2.70347

Vancouver

Baqleh AS, Alhouri N. Clinical Accuracy of an Intraoral Scanner for Tooth Shade Selection Compared With a Spectrophotometer and Visual Matching: A Comparative Study. Clinical and Experimental Dental Research. 2026. doi:10.1002/cre2.70347

BibTeX
@article{Baqleh2026,
  author = {Baqleh, Amr Sameer and Alhouri, Nabil},
  title = {Clinical Accuracy of an Intraoral Scanner for Tooth Shade Selection Compared With a Spectrophotometer and Visual Matching: A Comparative Study},
  journal = {Clinical and Experimental Dental Research},
  year = {2026},
  doi = {10.1002/cre2.70347}
}
Kostenlos starten →