EUVIMEDEUVIMED entdecken
Vergleichsstudie

Comparative Study Evaluating a New Dose Optimization Software for Gamma Knife Treatment Planning: Comparison of 80 Challenging Treatment Plans

Lucie Hamáčková, Josef Novotný, Markéta Farníková, Roman Liščák, Gabriela Šimonová, Dušan Urgošík, Michal Schmitt

Stereotactic and Functional Neurosurgery · 2025

DOI: 10.1159/000546214 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Stereotactic and Functional Neurosurgery, 2025. Verfasst von Lucie Hamáčková und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Na Homolce Hospital, Prague, Czechia. Hauptthemen laut MEDLINE: Radiosurgery, Software, Radiotherapy Planning, Computer-Assisted, Brain Neoplasms. Die Arbeit nennt 13 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

In 2020, Elekta Instrument AB, Stockholm, released a new dose optimizer for Leksell GammaPlan, which includes the possibility of inverse planning. This study aimed to compare the new software with the previous manual version of treatment planning for stereotactic radiosurgery and evaluate its performance. Materials and Methods: Four types of diagnoses – vestibular schwannomas, pituitary adenomas, meningiomas, and single brain metastasis – along with 80 clinically approved challenging cases, were selected for testing the new software. Key parameters, including coverage, selectivity, target volume, and doses to critical structures, were collected and statistically analysed using a t test. These parameters were compared based on the Leksell Gamma Knife (LGK) Society standardization document for stereotactic radiosurgery, both for each diagnosis and for the entire dataset.

Results

The new software showed a clear advantage, particularly in sparing critical structures while maintaining or improving treatment plan conformity. Doses to critical structures such as the optic nerve, brainstem, cochlea, and pituitary gland decreased by an average of 13% (0.76 Gy), 7% (0.52 Gy), 7% (0.2 Gy), and 14% (1.04 Gy), respectively, reducing toxicity. Other plan parameters also showed significant improvements, except for the gradient index. Selectivity improved by 11% (0.03), the Shaw Conformity Index improved by 10% (0.1), and coverage improved by 0.01. Additionally, treatment time was reduced by 10% enhancing patient comfort.

Conclusion

Overall, LGK Lightning is faster and produces treatment plans with superior parameters compared to manual planning.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgerySoftwareRadiotherapy Planning, Computer-AssistedBrain NeoplasmsHumansPituitary NeoplasmsRadiotherapy DosageNeuroma, AcousticMeningiomaMeningeal NeoplasmsMaleFemale

Stichwörter der Autor:innen

Dose Optimizer LightningLeksell Gamma KnifeLeksell GammaPlanTreatment planning

Autor:innen und Institutionen

  • Department of Medical Physics, Na Homolce Hospital, Prague, Czechia.
  • Department of Dosimetry and Application of Ionizing Radiation, Czech Technical University in Prague, Prague, Czechia.
  • Department of Stereotactic and Radiation Neurosurgery, Na Homolce Hospital, Prague, Czechia.

Zitieren

APA

Hamáčková, L., Novotný, J., Farníková, M., Liščák, R., Šimonová, G., Urgošík, D., & Schmitt, M. (2025). Comparative Study Evaluating a New Dose Optimization Software for Gamma Knife Treatment Planning: Comparison of 80 Challenging Treatment Plans. Stereotactic and Functional Neurosurgery. https://doi.org/10.1159/000546214

Vancouver

Hamáčková L, Novotný J, Farníková M, Liščák R, Šimonová G, Urgošík D, et al. Comparative Study Evaluating a New Dose Optimization Software for Gamma Knife Treatment Planning: Comparison of 80 Challenging Treatment Plans. Stereotactic and Functional Neurosurgery. 2025. doi:10.1159/000546214

BibTeX
@article{Hamkov2025,
  author = {Hamáčková, Lucie and Novotný, Josef and Farníková, Markéta and Liščák, Roman and Šimonová, Gabriela and Urgošík, Dušan and Schmitt, Michal},
  title = {Comparative Study Evaluating a New Dose Optimization Software for Gamma Knife Treatment Planning: Comparison of 80 Challenging Treatment Plans},
  journal = {Stereotactic and Functional Neurosurgery},
  year = {2025},
  doi = {10.1159/000546214}
}
Kostenlos starten →