Accuracy of injection pressure measurement at peripheral nerves using high-resolution 40 MHz ultrasound in an anesthetized porcine model
Regional Anesthesia & Pain Medicine · 2023-10 · 48 (10) · 501-507
DOI: 10.1136/rapm-2022-104282 ↗Kurz gefasst
Fachartikel in der Zeitschrift Regional Anesthesia & Pain Medicine, online erschienen am 23.02.2023. Verfasst von Graeme A McLeod und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anaesthesia, Ninewells Hospital, Dundee, UK g.a.mcleod@dundee.ac.uk. Hauptthemen laut MEDLINE: Nerve Block. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 29 Literaturangaben. Gefördert von: National Institute of Academic Anaesthesia.
Abstract (Originalsprache)
Background
Fluid injection pressure measurement is promoted as a marker of needle tip position that discriminates between tissue layers. However, clinical ultrasound has insufficient resolution to identify the exact position of the needle tip. Our primary objective was to use 40 MHz ultrasound in anesthetized pigs in order to precisely locate the tip of the needle and measure opening injection pressure in muscle, at epineurium and in subepineurium.
Methods
We surgically exposed the axillae of four anesthetized pigs. Two operators placed a 40 MHz ultrasound transducer over the pectoral muscle and imaged axillary, median and radial nerves. Injections (0.5 mL) were randomized to in-plane and out-of-plane needle trajectories and flow rates of 1, 6 and 12 mL/min.
Results
We identified 541 fascicles in 23 nerves. The ratio of fascicle area to nerve area remained constant at ~0.30 for all nerves. Axillary nerves were smaller than median and radial nerves, difference in diameter (95% CI) 1.61 (0.87 to 2.36) mm, p
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Anaesthesia, Ninewells Hospital, Dundee, UK g.a.mcleod@dundee.ac.uk.
- Institute of Academic Anaesthesia, University of Dundee, Dundee, UK.
- University of Dundee School of Medicine, Dundee, UK.
- Department of Anaesthesia, NHS Tayside, Dundee, UK.
- Department of Anesthesiology, CEU San Pablo University Faculty of Medicine, Madrid, Spain.
ORCID-Profile: Graeme A McLeod, Fiona Watson, Miguel Angel Reina
Förderung
- National Institute of Academic Anaesthesia
Zitieren
APA
McLeod, G. A., Cowie, A., Sadler, A., Watson, F., Wasik, P., & Reina, M. A. (2023). Accuracy of injection pressure measurement at peripheral nerves using high-resolution 40 MHz ultrasound in an anesthetized porcine model. Regional Anesthesia & Pain Medicine, 48(10), 501-507. https://doi.org/10.1136/rapm-2022-104282
Vancouver
McLeod GA, Cowie A, Sadler A, Watson F, Wasik P, Reina MA. Accuracy of injection pressure measurement at peripheral nerves using high-resolution 40 MHz ultrasound in an anesthetized porcine model. Regional Anesthesia & Pain Medicine. 2023;48(10):501-507. doi:10.1136/rapm-2022-104282
BibTeX
@article{McLeod2023,
author = {McLeod, Graeme A and Cowie, Alistair and Sadler, Amy and Watson, Fiona and Wasik, Paul and Reina, Miguel Angel},
title = {Accuracy of injection pressure measurement at peripheral nerves using high-resolution 40 MHz ultrasound in an anesthetized porcine model},
journal = {Regional Anesthesia & Pain Medicine},
year = {2023},
volume = {48},
number = {10},
pages = {501--507},
doi = {10.1136/rapm-2022-104282}
}